What to do if... MultiLab 4010-2
90 ba76142e01 05/2013
No stable measured
value
Obviously incorrect
measured values
13.2 Dissolved oxygen
Error message,
OFL
Error message,
Error
Cause Remedy
IDS pH sensor:
Diaphragm contaminated Clean diaphragm
Membrane contaminated Clean membrane
Test sample:
pH value not stable Measure with air excluded if
necessary
Temperature not stable Adjust temperature if necessary
IDS pH sensor + test sample:
Conductivity too low Use suitable IDS pH sensor
Temperature too high Use suitable IDS pH sensor
Organic liquids Use suitable IDS pH sensor
Cause Remedy
IDS pH sensor:
IDS pH sensor unsuitable Use suitable IDS sensor
Temperature difference between
buffer and test sample too great
Adjust temperature of buffer or
sample solutions
Measurement procedure not suit-
able
Follow special procedure
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause Remedy
Measured value outside the measur-
ing range
Select a suitable measuring
medium
Cause Remedy
Sensor contaminated Clean the sensor