pH ProLab 3000
70
ba75645e03 12/2012
Criteria With identical measurement conditions, the following applies:
Note
You can terminate prematurely the Stability control function with
<OK> manually at any time. If the Stability control function is termi-
nated prematurely, the current measurement data is not downloaded
to the interface.
5 Using <SC> or <MODE>, release the frozen measured value.
The [SC] status display disappears.
Measured param-
eter
Reproducibility Response time
pH value Better than 0.01 > 15 seconds