What to do if... ProLab 3000
176
ba75645e03 12/2012
Warning [TempErr]
Warning [ISEErr]
Burette does not dose
15.3 General errors
Meter does not react to
keystroke
Time is lost
Cause Remedy
Temperature difference
between measurement and cali-
bration greater than 2 K.
Adjust the temperature of
the test sample
Cause Remedy
Electrode voltage outside cali-
brated range
Recalibrate
Cause Remedy
Address of the burette incor-
rectly set
Enter the address of the
burette on the meter
(see page 99)
Enter the address of the
burette on the burette
(see operating manual of the
burette)
RS232 interface of the burette
incorrectly set
Configure the RS232 inter-
face of the burette
(see page 29 and operating
manual of the burette)
Cause Remedy
Operating condition undefined
or EMC load unallowed
Processor reset:
Press the <On/Off> and
<SC> key simultaneously.
Cause Remedy
Buffer batteries empty Exchange buffer batteries
(see page 171)