What to do if... ProLab 3000
174
ba75645e03 12/2012
No stable measured
value
Obviously incorrect
measured values
Electrode broken Replace the electrode
Buffer solutions:
Incorrect buffer solutions Select the suitable buffer set
for the buffers used
or
Replace the buffer solutions
Buffer solutions too old Use only once.
Note the shelf life
Buffer solutions depleted Replace the buffer solutions
Cause Remedy
Cause Remedy
pH electrode:
Diaphragm contaminated Clean diaphragm
Membrane contaminated Clean membrane
Test sample:
pH value not stable Measure with air excluded if
necessary
Temperature not stable Adjust temperature if neces-
sary
Electrode + test sample:
Conductivity too low Use a suitable electrode
Temperature too high Use a suitable electrode
Organic liquids Use a suitable electrode
Cause Remedy
pH electrode:
Not connected Connect the electrode
Cable broken Replace the electrode
pH electrode unsuitable Use a suitable electrode
Temperature difference
between buffer and test sam-
ple too high
Adjust temperature of buffer
or sample solutions