Technical Data ProLab 3000
182
ba75645e03 12/2012
16.2 Measuring ranges, resolution, accuracy
16.2.1 Temperature
16.2.2 pH/ORP
Measuring ranges,
resolution
Variable Measuring range Resolution
T [°C] - 35.0 ... + 150.0 0.1
Accuracy (± 1 digit) Variable Accuracy Temperature of the
test sample
T [°C] ± 0.1 - 35.0 ... + 150.0
Manual
temperature input
Variable Range Increment
T
manual
[°C] -35 ... + 150 1
Measuring ranges,
resolution
Variable Measuring range Resolution
pH - 2.000 ... + 20.000
- 2.00 ... + 20.00
- 2.0 ... + 20.0
0.001
0.01
0.1
U [mV] - 2200.0 ... + 2200.0
- 2200 ... + 2200
0.1
1
Accuracy (± 1 digit) Variable Accuracy Temperature of the test
sample
pH ± 0.002
± 0.01
± 0.1
+ 15 °C ... + 35 °C
U [mV] / range
- 2200.0 ... + 2200.0 ± 0.1 + 15 °C ... + 35 °C
- 2200 ... + 2200 ± 1 + 15 °C ... + 35 °C