Xylem ProLab 3000 Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Ferramentas Xylem ProLab 3000. Xylem ProLab 3000 User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 202
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes

Resumo do Conteúdo

Página 1 - ProLab 3000

ba75645e03 12/2012DRAFT December 10, 2012 2:50 pmProLab 3000 Operating manualpH/ISE meter with automatic sensor recognition and electronic access cont

Página 2 - © 2012, SI Analytics GmbH

Contents ProLab 300010ba75645e03 12/2012

Página 3 - DÉCLARATION DE CONFORMITÉ

Ion concentration ProLab 3000100ba75645e03 12/2012– Address of TISAB/ISA : (1...3)– Address of standard/sample : (1...3)– (Address of blank value:) (1

Página 4

ProLab 3000 Ion concentration101ba75645e03 12/20129.3 CalibrationWhy calibrate? Ion-selective electrodes age and are temperature-dependent. This chang

Página 5 - ProLab 3000 - Contents

Ion concentration ProLab 3000102ba75645e03 12/2012ISE Cal This is the conventional two-point to nine-point calibration proce-dure that uses 2 to 9 fre

Página 6

ProLab 3000 Ion concentration103ba75645e03 12/2012Calibration evaluation After calibrating, the meter automatically evaluates the calibration.Display

Página 7 - ProLab 3000 Contents

Ion concentration ProLab 3000104ba75645e03 12/20129.3.1 Carrying out calibrationUse two to nine standard solutions in ascending or descending order.Pr

Página 8

ProLab 3000 Ion concentration105ba75645e03 12/20123 When measuring without temperature sensor: Measure the temperature of the standard solution using

Página 9

Ion concentration ProLab 3000106ba75645e03 12/2012Continuing with two-point calibration9 Wait for the measurement with stability control to end.orTake

Página 10 - Contents ProLab 3000

ProLab 3000 Ion concentration107ba75645e03 12/2012Continuing with three-point to nine-point calibrationRepeat the steps 9 to 16 in the same way with t

Página 11 - 1Overview

Ion concentration ProLab 3000108ba75645e03 12/20129.3.2 Calibration dataIn the Calibration menu, you can view the calibration data of ISE cali-bration

Página 12 - 1.2 Keypad

ProLab 3000 Ion concentration109ba75645e03 12/20129.4 Blank value correctionIf the ion concentration in the sample is so small that it is not in the l

Página 13 - ProLab 3000 Overview

ProLab 3000 Overview11ba75645e03 12/20121Overview1.1 General featuresThe ProLab 3000 precision pH meter enables you to perform pH and ORP measurements

Página 14 - 1.3 Display

Ion concentration ProLab 3000110ba75645e03 12/20129.5 Reference measurementA reference measurement enables you to assign any concentration value to th

Página 15 - 1.4 Socket field

ProLab 3000 Ion concentration111ba75645e03 12/20129.6 Measuring with increment procedure (methods)9.6.1 Selecting the measuring methodThe following me

Página 16 - Overview ProLab 3000

Ion concentration ProLab 3000112ba75645e03 12/20129.6.2 Standard additionIn the Standard addition procedure, a known amount of standard solu-tion is a

Página 17

ProLab 3000 Ion concentration113ba75645e03 12/20126 In the Sample register, enter the values for the sample volume (Sample volume) and the volume of t

Página 18

Ion concentration ProLab 3000114ba75645e03 12/201212 In the Standard register, enter the values for Standard volume and Standard concentration if nece

Página 19

ProLab 3000 Ion concentration115ba75645e03 12/201216 Change the displayed unit of the measured value with [Unit] if necessary.17 If required, display

Página 20 - 1.6 Electronic access control

Ion concentration ProLab 3000116ba75645e03 12/20129.6.3 Double standard additionIn the Double standard addition procedure, a known amount of stan-dard

Página 21 - 2 Safety

ProLab 3000 Ion concentration117ba75645e03 12/20125 Thoroughly rinse the electrode with deionized water.6 Immerse the electrode in the sample.7 Start

Página 22 - Safety ProLab 3000

Ion concentration ProLab 3000118ba75645e03 12/201210 Start the measurement with [Continue].The measured value is checked for stability (Stability cont

Página 23 - 3 Commissioning

ProLab 3000 Ion concentration119ba75645e03 12/201215 Change the displayed unit of the measured value with [Unit] if necessary.16 If required, display

Página 24 - 3.2 Initial commissioning

Overview ProLab 300012ba75645e03 12/20121.2 KeypadKeys with dedicated functionsKeys with dedicated functions are labeled on the meter. In this operat

Página 25 - ProLab 3000 Commissioning

Ion concentration ProLab 3000120ba75645e03 12/20129.6.4 Standard subtractionIn the "Standard Subtraction" procedure, a known amount of stand

Página 26 - 3.6 Connecting sensors

ProLab 3000 Ion concentration121ba75645e03 12/20126 In the Sample register, enter the values for the sample volume (Sample volume) and the volume of t

Página 27

Ion concentration ProLab 3000122ba75645e03 12/201213 In the Standard register, enter the values for Standard volume and Standard concentration if nece

Página 28 - Commissioning ProLab 3000

ProLab 3000 Ion concentration123ba75645e03 12/201217 Change the displayed unit of the measured value with [Unit] if necessary.18 If required, display

Página 29

Ion concentration ProLab 3000124ba75645e03 12/20129.6.5 Sample additionIn the Sample addition procedure, a known amount of test sample is added to the

Página 30

ProLab 3000 Ion concentration125ba75645e03 12/20126 In the Standard register, enter the values for Standard vol-umeStandard concentration and ISA/TISA

Página 31

Ion concentration ProLab 3000126ba75645e03 12/201212 In the Sample register, enter the values for the sample volume (Sample volume) if necessary.13 Us

Página 32

ProLab 3000 Ion concentration127ba75645e03 12/201216 Change the displayed unit of the measured value with [Unit] if necessary.17 If required, display

Página 33 - 4 Operating principles

Ion concentration ProLab 3000128ba75645e03 12/20129.6.6 Sample subtractionIn the Sample subtraction procedure, a known amount of test sample is added

Página 34

ProLab 3000 Ion concentration129ba75645e03 12/20126 In the Standard register, enter the values for Standard vol-umeStandard concentration and ISA/TISA

Página 35

ProLab 3000 Overview13ba75645e03 12/2012Alphanumeric keypad The keys of the alphanumeric keypad are used to enter numerals (<0...9>), letters (&

Página 36 - 4.2 File system

Ion concentration ProLab 3000130ba75645e03 12/201213 In the Sample register, enter the value for the sample volume (Sample volume) if necessary.14 Use

Página 37

ProLab 3000 Ion concentration131ba75645e03 12/201217 Change the displayed unit of the measured value with [Unit] if necessary.18 If required, display

Página 38

Ion concentration ProLab 3000132ba75645e03 12/20129.6.7 Blank value addition (Standard addition with blank value correction)In the "Standard add

Página 39 - 4.4 Navigation

ProLab 3000 Ion concentration133ba75645e03 12/2012Measuring6 In the Sample register, enter the following values: the volume of the sample (Sample vol

Página 40

Ion concentration ProLab 3000134ba75645e03 12/20129 Immerse the electrode in the sample that was supplemented with blank value solution.10 Start the m

Página 41

ProLab 3000 Ion concentration135ba75645e03 12/201214 Start the measurement with [Continue].The measured value is checked for stability (Stability cont

Página 42

Ion concentration ProLab 3000136ba75645e03 12/2012

Página 43

ProLab 3000 Memory137ba75645e03 12/201210 MemoryWith the ProLab 3000 software you can store and recall all important measurement and configuration dat

Página 44

Memory ProLab 3000138ba75645e03 12/201210.1.1 Creating and selecting a file for manually stored measurement datasetsCreating a file for manually store

Página 45

ProLab 3000 Memory139ba75645e03 12/201210.1.2 Storing measurement datasets manuallyMeasurement datasets are stored in a file selected before (see page

Página 46

Overview ProLab 300014ba75645e03 12/20121.3 DisplayThe graphic color display displays the measurement data. The illumi-nation enables to read the disp

Página 47 - 5 Access to the meter

Memory ProLab 3000140ba75645e03 12/201210.1.3 Automatically storing measurement datasetsThe file in which measurement datasets should be automatically

Página 48

ProLab 3000 Memory141ba75645e03 12/2012 NoteThe progress of the automatic memory is indicated by a progress bar. The Automatic memory window can b

Página 49 - 5.3 Password for login

Memory ProLab 3000142ba75645e03 12/201210.1.4 Displaying and filtering stored measurement datasets1 Open the currently selected manual memory with <

Página 50 - 5.4 Lock

ProLab 3000 Memory143ba75645e03 12/201210.1.5 Erasing measurement datasets4 Activate the selected filter criteria with [OK].The [Filter] option field

Página 51

Memory ProLab 3000144ba75645e03 12/201210.2 Calibration dataThe current calibration record is automatically stored after each valid calibration proced

Página 52

ProLab 3000 Memory145ba75645e03 12/201210.2.2 Managing the calibration dataThe calibration data is stored in the current calibration record (menu, Cal

Página 53

Memory ProLab 3000146ba75645e03 12/201210.3 Recorder dataInformation on the recording, storing, displaying and deleting of mea-sured value curves is g

Página 54

ProLab 3000 Recorder147ba75645e03 12/201211 RecorderThe recorder represents measured values graphically in a system of coordinates. The recorder can r

Página 55 - 5.6 Lost your electronic key?

Recorder ProLab 3000148ba75645e03 12/201211.1 Recording the measured value curveThe recorder records the course of the measured values curve and the c

Página 56 - 6 System functions

ProLab 3000 Recorder149ba75645e03 12/201211.2 Recording the curve of the measured temperature value If the recorder was started without recor

Página 57 - ProLab 3000 System functions

ProLab 3000 Overview15ba75645e03 12/20121.4 Socket fieldConnections:1 pH/ISE/ORP electrode (pH/U/ISE Probe 2)2 Reference electrode (Ref 2)3 Temperatur

Página 58 - 6.3 Selecting the country

Recorder ProLab 3000150ba75645e03 12/201211.3 Storing recorder dataThe recorder records the course of the measured values curve and the course of the

Página 59 - 6.4 Setting the date and time

ProLab 3000 Recorder151ba75645e03 12/201211.4 Displaying stored recorder dataOpening the offline recorderDisplaying a different recorder file1 Open th

Página 60 - System functions ProLab 3000

Recorder ProLab 3000152ba75645e03 12/201211.5 Settings for the recorderIn the Recorder settings dialog box, the appearance of your recorder is defined

Página 61 - 6.7 Temperature

ProLab 3000 Recorder153ba75645e03 12/201211.5.1 Defining the recording intervalThe Recording interval defines at which intervals the measured values a

Página 62

Recorder ProLab 3000154ba75645e03 12/20121 Open the recorder menu, Settings...Settings...The Recorder settings window pops up.2 Select the Measured v

Página 63

ProLab 3000 Recorder155ba75645e03 12/201211.5.3 Defining and activating limit valuesWith the limit values you define the measured values at which the

Página 64 - 6.9 Color assignment

Recorder ProLab 3000156ba75645e03 12/201211.5.4 Defining the colors for the recorder representation The appearance of the recorder is determined w

Página 65 - 6.10 Reset

ProLab 3000 Recorder157ba75645e03 12/201211.6 Changing the displayed axis interceptThe standard settings for the displayed axis intercepts are defined

Página 66 - 6.11 List of open windows

Recorder ProLab 3000158ba75645e03 12/2012Zooming an axis interceptThe axis intercepts of the recorder can be zoomed as follows (time axis and measured

Página 67 - 7.1 General information

ProLab 3000 Recorder159ba75645e03 12/201211.7 Displaying exact measurement data at a point (cursor)In the recorder window, you can display the exact m

Página 68 - 7.2 Measuring the pH value

Overview ProLab 300016ba75645e03 12/2012 CAUTIONOnly connect sensors to the meter that cannot return any voltag-es or currents that are not allowed (

Página 69 - ProLab 3000 pH

Recorder ProLab 3000160ba75645e03 12/2012

Página 70

ProLab 3000 Transmitting data161ba75645e03 12/201212 Transmitting data12.1 Printing current measurement dataTo which interface the measurement data sh

Página 71

Transmitting data ProLab 3000162ba75645e03 12/201212.2.2 USB-B interface (USB Device)Connect the interface to the PC via the supplied Z875 USB cable.

Página 72 - 7.3 Calibration

ProLab 3000 Transmitting data163ba75645e03 12/201212.2.3 USB-A interface (USB Host)Connect the USB-A interface (USB Host) of the ProLab 3000 to the pr

Página 73

Transmitting data ProLab 3000164ba75645e03 12/201212.3 Data backup12.3.1 Data backup by the administratorTo back up important data against being lost

Página 74

ProLab 3000 Transmitting data165ba75645e03 12/201212.3.2 Viewing backed up dataThe external USB medium is displayed within the directory of each user.

Página 75

Configurations ProLab 3000166ba75645e03 12/201213 ConfigurationsAs the administrator, you can create and assign a configuration to each user (see page

Página 76 - HAT TO DO IF... (page 173)

ProLab 3000 Configurations167ba75645e03 12/201213.1 Viewing a configurationA configuration viewer is available so the current configuration settings c

Página 77

Configurations ProLab 3000168ba75645e03 12/201213.2 Creating a new configurationWhen creating a configuration, the current settings of the meter for e

Página 78

ProLab 3000 Configurations169ba75645e03 12/201213.3 Loading a configurationAll users who can create a configuration can also load it.When loading a co

Página 79

ProLab 3000 Overview17ba75645e03 12/20121.5 Automatic sensor recognitionThe automatic sensor recognition function enables  to operate a sensor with d

Página 80

Configurations ProLab 3000170ba75645e03 12/201213.4 Deleting a configurationAll users who can create a configuration can also delete it in their direc

Página 81

ProLab 3000 Maintenance, cleaning, disposal171ba75645e03 12/201214 Maintenance, cleaning, disposal14.1 MaintenanceThe only maintenance activity requir

Página 82

Maintenance, cleaning, disposal ProLab 3000172ba75645e03 12/201214.2 CleaningOccasionally wipe the outside of the meter with a damp, lint-free cloth.

Página 83

ProLab 3000 What to do if...173ba75645e03 12/201215 What to do if...15.1 pH and ORP measurementError message Err1Error message Err2Error message Err4E

Página 84 - Note

What to do if... ProLab 3000174ba75645e03 12/2012No stable measured value Obviously incorrect measured values– Electrode broken – Replace the electr

Página 85

ProLab 3000 What to do if...175ba75645e03 12/2012CalClock on red background15.2 Ion selective measurementError message Err1 Obviously incorrect measur

Página 86

What to do if... ProLab 3000176ba75645e03 12/2012Warning [TempErr] Warning [ISEErr] Burette does not dose 15.3 General errorsMeter does not react t

Página 87 - 8 ORP voltage

ProLab 3000 What to do if...177ba75645e03 12/2012

Página 88 - 8.2 Measuring the ORP

What to do if... ProLab 3000178ba75645e03 12/2012

Página 89 - ProLab 3000 ORP voltage

ProLab 3000 Technical Data179ba75645e03 12/201216 Technical Data16.1 General dataDimensions approx. 240 x 280 x 70 mmWeight approx. 2.5 kg (without po

Página 90 - ORP voltage ProLab 3000

Overview ProLab 300018ba75645e03 12/20121.5.1 ID sensorsSI Analytics GmbH ID sensors support the automatic sensor recogni-tion function. Their sensor

Página 91

Technical Data ProLab 3000180ba75645e03 12/2012RS232 serial interface Connection of the PC cable (Z390) or printer cable (Z393)Baud rate adjustable:

Página 92

ProLab 3000 Technical Data181ba75645e03 12/2012Instrument safety EC guideline 2006/95/ECEN 61010-1 class AANSI/UL 61010-1CAN/CSA-C22.2 No. 61010-1Rad

Página 93

Technical Data ProLab 3000182ba75645e03 12/201216.2 Measuring ranges, resolution, accuracy16.2.1 Temperature16.2.2 pH/ORPMeasuring ranges, resolutionV

Página 94

ProLab 3000 Technical Data183ba75645e03 12/201216.2.3 ISEMeasuring ranges, resolutionVariable Measuring range ResolutionISE [mg/l, %, ppm, mg/kg, µm

Página 95 - 9 Ion concentration

Technical Data ProLab 3000184ba75645e03 12/2012

Página 96

ProLab 3000 Lists185ba75645e03 12/201217 ListsThis chapter provides additional information and orientation aids.Abbreviations The list of abbreviation

Página 97

Lists ProLab 3000186ba75645e03 12/2012Abbreviations°C Temperature unit, degrees Celsius°F Temperature unit, degrees FahrenheitLoBat Batteries almost e

Página 98

ProLab 3000 Lists187ba75645e03 12/2012Status display indicators[SC] Measured value (Hold) or Measurement with stability control[TP] Temperature senso

Página 99

Lists ProLab 3000188ba75645e03 12/2012GlossaryAdjusting To manipulate a measuring system so that the relevant value (e. g. the displayed value) differ

Página 100 - Ion concentration ProLab 3000

ProLab 3000 Lists189ba75645e03 12/2012ORP voltage The ORP is caused by oxidizing or reducing substances dissolved in water if these substances become

Página 101 - 9.3 Calibration

ProLab 3000 Overview19ba75645e03 12/20121.5.2 Sensor data from ID sensorsID sensors transmit the following sensor data to the meter: SENSOR ID– Senso

Página 102

Lists ProLab 3000190ba75645e03 12/2012Trademarks usedTrademark Owner of the trademarkMerck Merck KGaASCHOTT Schott Glas AGMettler Toledo Mettler Toled

Página 103 - ProLab 3000 Ion concentration

ProLab 3000 Lists 191Indexba75645e03 12/2012AAuthorized use ...21BBattery compartment ...

Página 104

Index ProLab 3000192ba75645e03 12/2012SSafe operation ... 22Safety ...

Página 105

ProLab 3000 Appendix193ba75645e03 12/2012AppendixA.18 Firmware updateGeneral information With the "Update_ProLab3000_4000" program and a PC

Página 106

Appendix ProLab 3000194ba75645e03 12/2012Program installationProgram start Firmware updateAfter switching the meter off and on you can check whether t

Página 107

ProLab 3000 Appendix195ba75645e03 12/2012A.19 MenusA.19.1 Main menuFile MemoryFile – New configuration...– Select configuration...– View configuration

Página 108

Appendix ProLab 3000196ba75645e03 12/2012SystemWindowUser Help System– Data output...– Baud– Output format (ASCII, CSV)–– Header– Autom. printout...–

Página 109 - 9.4 Blank value correction

ProLab 3000 Appendix197ba75645e03 12/2012A.19.2 pH sensor menuMeasuringCalibrationMeasuring–pH resolution–High– Medium– Low– Alternative TP– Setting o

Página 110 - 9.5 Reference measurement

Appendix ProLab 3000198ba75645e03 12/2012A.19.3 U sensor menuMeasuringA.19.4 dU sensor menuMeasuringMeasuring– High resolution– Alternative TP– Settin

Página 111

ProLab 3000 Appendix199ba75645e03 12/2012A.19.5 ISE sensor menuMeasuringCalibrationMeasuring– ISE settings–Units– Ion type– Blank value correction– Re

Página 112

Accuracy when going to pressThe use of advanced technology and the high quality standard of our instruments are the result of a continuous development

Página 113

Overview ProLab 300020ba75645e03 12/20121.6 Electronic access controlThe ProLab 3000 always documents measurement and calibration data together with a

Página 114

Appendix ProLab 3000200ba75645e03 12/2012

Página 116

SI Analytics GmbHPostfach 2443D-55014 MainzHattenbergstr. 10D-55122 MainzEmail: support@Telefon +49 (0) 61 31/66 5111Telefax +49 (0) 61 31/66 5001si-a

Página 117

ProLab 3000 Safety21ba75645e03 12/20122 SafetyThis operating manual contains basic instructions that you must follow during the commissioning, operati

Página 118

Safety ProLab 300022ba75645e03 12/20122.2 General safety instructionsThis instrument is built and inspected according to the relevant guide-lines and

Página 119

ProLab 3000 Commissioning23ba75645e03 12/20123 Commissioning3.1 Scope of delivery ProLab 3000 laboratory meter Power pack 4 batteries 1.5 V Micro t

Página 120

Commissioning ProLab 300024ba75645e03 12/20123.2 Initial commissioningPerform the following activities: Insert the batteries (see page 171) Connect

Página 121

ProLab 3000 Commissioning25ba75645e03 12/20123.4 Switching on the ProLab 3000The ProLab 3000 indicates the Standby display.1 Switch the meter on with

Página 122

Commissioning ProLab 300026ba75645e03 12/20123.5 Connecting the USB mouseYou can connect the USB mouse included in the scope of delivery to the USB-A

Página 123

ProLab 3000 Commissioning27ba75645e03 12/20123.6.1 Connecting an ID sensorID sensor with temperature sensorID sensor without temperature sensor1 Conne

Página 124

Commissioning ProLab 300028ba75645e03 12/20123.6.2 Connecting a non ID sensor3.7 Connecting optional accessoriesThe ProLab 3000 has several interfaces

Página 125

ProLab 3000 Commissioning29ba75645e03 12/20123.7.1 RS232 interface (serial port)PC, printer and burettes can be connected via the RS232 interface.Data

Página 126

ProLab 30003ba75645e03 12/2012KONFORMITÄTSERKLÄRUNGDECLARATION OF CONFORMITYDÉCLARATION DE CONFORMITÉWir erklären in alleiniger Verantwortung, dass da

Página 127

Commissioning ProLab 300030ba75645e03 12/2012

Página 128

ProLab 3000 Commissioning31ba75645e03 12/20123.7.2 USB-B interface (USB Device)You can connect a PC to the meter via the USB-B interface(USB Device).

Página 129

Commissioning ProLab 300032ba75645e03 12/20123.7.3 USB-A interface (USB Host)You can connect the following devices to the USB-A interface (USB Host) a

Página 130

ProLab 3000 Operating principles33ba75645e03 12/20124 Operating principlesThis section contains basic information on the operation of the ProLab 3000.

Página 131

Operating principles ProLab 300034ba75645e03 12/2012Display Depending on the operating situation, the color display shows mea-surement data, calibrati

Página 132

ProLab 3000 Operating principles35ba75645e03 12/2012Channel (2) In the measured value display, for each connected sensor you can see a channel with sp

Página 133

Operating principles ProLab 300036ba75645e03 12/20124.2 File systemFiles With the ProLab 3000 you can store data in files just like using a PC. Differ

Página 134

ProLab 3000 Operating principles37ba75645e03 12/20124.3 Entry of numerals, letters and charactersNumerals, letters, punctuation marks and special char

Página 135

Operating principles ProLab 300038ba75645e03 12/2012Mouse operation:With the mouse, you have in all entry fields the additional option of showing a ke

Página 136

ProLab 3000 Operating principles39ba75645e03 12/20124.4 NavigationThe dialog with the meter takes place through menus and dialog boxes as usual with a

Página 137 - 10 Memory

ProLab 30004ba75645e03 12/2012

Página 138 - Memory ProLab 3000

Operating principles ProLab 300040ba75645e03 12/20124.4.1 Navigation in the measured value displayThe display shows, for example, the measurement data

Página 139 - ProLab 3000 Memory

ProLab 3000 Operating principles41ba75645e03 12/20124.4.2 Navigation in menusA menu item contains submenus, executes a function or opens a dialog box.

Página 140

Operating principles ProLab 300042ba75645e03 12/2012Mouse operation:5 Highlight a menu item with <><> and <><>.The highlig

Página 141

ProLab 3000 Operating principles43ba75645e03 12/20124.4.3 Navigation in dialog boxesDialog boxes contain further subelements such as register cards, s

Página 142

Operating principles ProLab 300044ba75645e03 12/20124.4.4 Navigation in the file selection dialog boxThe file selection dialog is opened if a file is

Página 143

ProLab 3000 Operating principles45ba75645e03 12/2012Operating example: Navigation and settings in file selection dialogsMouse operation:1 Highlight an

Página 144 - 10.2 Calibration data

Operating principles ProLab 300046ba75645e03 12/2012

Página 145

ProLab 3000 Access to the meter47ba75645e03 12/20125 Access to the meter5.1 Switch the meter on and offSwitching onSwitching off5.2 Login as a userAft

Página 146 - 10.4 Configuration data

Access to the meter ProLab 300048ba75645e03 12/2012Login with electronic keyThe Login dialog box displays the instruction: Place the electronic key o

Página 147 - 11 Recorder

ProLab 3000 Access to the meter49ba75645e03 12/20125.3 Password for loginIf the used of a password for login to the meter was set by the admin-istrato

Página 148 - Recorder ProLab 3000

ProLab 3000 Contents5ba75645e03 12/2012ProLab 3000 - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111.

Página 149 - ProLab 3000 Recorder

Access to the meter ProLab 300050ba75645e03 12/20125.4 LockDuring operation (e.g. if the automatic storing function is active), the activated lock pre

Página 150 - 11.3 Storing recorder data

ProLab 3000 Access to the meter51ba75645e03 12/20125.5 Access control and user rightsYou as the administrator define the type of access (assess contro

Página 151

Access to the meter ProLab 300052ba75645e03 12/20125.5.1 Access controlThe administrator defines the type of access to the meter. The setting applies

Página 152

ProLab 3000 Access to the meter53ba75645e03 12/20125.5.2 User management and assigning user rightsThe ProLab 3000 has basic user management functions.

Página 153

Access to the meter ProLab 300054ba75645e03 12/2012Adding a new userAssigning a configuration to a userErasing a registered user1 Enter a new user nam

Página 154

ProLab 3000 Access to the meter55ba75645e03 12/2012Erasing the password for a registered userHiding / displaying a userExiting the user management5.6

Página 155

System functions ProLab 300056ba75645e03 12/20126 System functionsSystem functions are all functions that are independent of the probe. 6.1 Overview:

Página 156

ProLab 3000 System functions57ba75645e03 12/2012– Signal On () Switches on or off the acoustic signal. The signal sounds e.g. when a stable measured

Página 157

System functions ProLab 300058ba75645e03 12/20126.2 Selecting the languageMenus can be displayed in different languages. The language is selected in t

Página 158

ProLab 3000 System functions59ba75645e03 12/20126.4 Setting the date and timeThe data and time are displayed in the local format, according to the Cou

Página 159

Contents ProLab 30006ba75645e03 12/20125 Access to the meter . . . . . . . . . . . . . . . . . . . . . . . . . . . 475.1 Switch the meter on and off

Página 160

System functions ProLab 300060ba75645e03 12/20126.5 Selecting the channels for measured value displayIn the Window / Channels... you can define which

Página 161 - 12 Transmitting data

ProLab 3000 System functions61ba75645e03 12/20126.7 TemperatureThe temperature affects the measurement results. You can select the unit of the tempera

Página 162 - Transmitting data ProLab 3000

System functions ProLab 300062ba75645e03 12/20126.7.2 Temperature measurementFor reproducible measurements it is essential to determine the corre-spon

Página 163 - ProLab 3000 Transmitting data

ProLab 3000 System functions63ba75645e03 12/20126.8 Automatic stability controlThe Autom. stability control function continuously checks the stability

Página 164 - 12.3 Data backup

System functions ProLab 300064ba75645e03 12/20126.9 Color assignmentFor easy differentiation, the following conditions are highlighted by a background

Página 165

ProLab 3000 System functions65ba75645e03 12/20126.10 ResetYou can reset (initialize) all sensor settings and sensor-independent settings separately fr

Página 166 - 13 Configurations

System functions ProLab 300066ba75645e03 12/20126.11 List of open windowsThe Window menu item in the main menu provides an overview of the open window

Página 167 - 13.1 Viewing a configuration

ProLab 3000 pH67ba75645e03 12/20127pH7.1 General informationTemperature measurementFor reproducible pH measurements, it is essential to measure the te

Página 168 - Configurations ProLab 3000

pH ProLab 300068ba75645e03 12/20127.2 Measuring the pH value7.2.1 Preparatory activitiesPerform the following preparatory activities when you want to

Página 169 - 13.3 Loading a configuration

ProLab 3000 pH69ba75645e03 12/20127.2.2 MeasuringStability control The Stability control function checks the stability of the measurement signal. The

Página 170 - 13.4 Deleting a configuration

ProLab 3000 Contents7ba75645e03 12/20128 ORP voltage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 878.1 General information. .

Página 171 - 14.1 Maintenance

pH ProLab 300070ba75645e03 12/2012Criteria With identical measurement conditions, the following applies: NoteYou can terminate prematurely the Sta

Página 172 - 14.3 Disposal

ProLab 3000 pH71ba75645e03 12/20127.2.3 Measurement settingsAll settings for pH measurements are in the Measuring menu.Menu item Setting DescriptionMe

Página 173 - 15 What to do if

pH ProLab 300072ba75645e03 12/20127.3 CalibrationWhy calibrate? pH electrodes age. This changes the zero point (asymmetry) and slope of the pH electro

Página 174 - What to do if... ProLab 3000

ProLab 3000 pH73ba75645e03 12/20128 DIN 19267 * 1,0904,6506,7909,23025 °C9 Mettler EU 1,6794,0037,00210,01325 °C10 Mettler US * 1,9954,0057,0029,20825

Página 175 - ± 2 °C temperature differ

pH ProLab 300074ba75645e03 12/2012* Brand names or trade names are trademarks of their respective owners protected by law (see page 190).Calibration p

Página 176 - 15.3 General errors

ProLab 3000 pH75ba75645e03 12/2012Stability control The calibration procedure automatically activates the stability control function. The current meas

Página 177 - ProLab 3000 What to do if

pH ProLab 300076ba75645e03 12/2012Calibration evaluation After calibrating, the meter automatically evaluates the calibration. The zero point and slop

Página 178

ProLab 3000 pH77ba75645e03 12/20127.3.1 Carrying out a calibration procedure (example: AutoCal DIN) Make sure the Calibration / pH calibratio

Página 179 - 16 Technical Data

pH ProLab 300078ba75645e03 12/20121 Recognized nominal buffer value (referring to 25 °C)2 Current electrode voltage3 Current temperature value4 Select

Página 180 - Technical Data ProLab 3000

ProLab 3000 pH79ba75645e03 12/2012Continuing with two-point calibration (Buffer: DIN)9 Using [OK], confirm the Finish with 1-point calibration infotex

Página 181 - ProLab 3000 Technical Data

Contents ProLab 30008ba75645e03 12/2012temperature value . . . . . . . . . . . . . . . . . . . . . . . . 14911.3 Storing recorder data . . .

Página 182

pH ProLab 300080ba75645e03 12/2012Continuing with three-point to five-point calibration (Buffer ) DIN) NoteA calibration line is determined for the

Página 183

ProLab 3000 pH81ba75645e03 12/20127.3.2 Carrying out a calibration procedure (example: VariCal)Make sure the Calibration / pH calibration type / VariC

Página 184

pH ProLab 300082ba75645e03 12/20121 Recognized nominal buffer value (referring to 25 °C)2 Current electrode voltage3 Current temperature value4 Select

Página 185 - 17 Lists

ProLab 3000 pH83ba75645e03 12/2012Continuing with two-point calibration8 Using [OK], confirm the Finish with 1-point calibration infotext and take ove

Página 186 - Lists ProLab 3000

pH ProLab 300084ba75645e03 12/2012 NoteA calibration line is determined for the two-point calibration.

Página 187 - ProLab 3000 Lists

ProLab 3000 pH85ba75645e03 12/20127.3.3 Calibration settings and calibration dataIn the Calibration menu, you can  set the calibration interval,  de

Página 188

pH ProLab 300086ba75645e03 12/20127.3.4 Calibration intervalThe calibration interval and calibration evaluation are indicated on the display as the Ca

Página 189

ProLab 3000 ORP voltage87ba75645e03 12/20128 ORP voltage8.1 General informationYou can measure the following variables: ORP [mV] Relative ORP [mV]Te

Página 190

ORP voltage ProLab 300088ba75645e03 12/20128.2 Measuring the ORP8.2.1 Preparatory activitiesPerform the following preparatory activities when you want

Página 191

ProLab 3000 ORP voltage89ba75645e03 12/20128.2.2 MeasuringStability control TheStability control function checks the stability of the measurement sign

Página 192 - Index ProLab 3000

ProLab 3000 Contents9ba75645e03 12/201217 Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185Appendix . . . . .

Página 193 - Appendix

ORP voltage ProLab 300090ba75645e03 12/2012Criteria With identical measurement conditions, the following applies:4With <OK> activate the Stabili

Página 194 - Appendix ProLab 3000

ProLab 3000 ORP voltage91ba75645e03 12/20128.2.3 Measuring the relative ORPTo measure the difference of the ORPs of two solutions it is best to use tw

Página 195 - A.19 Menus

ORP voltage ProLab 300092ba75645e03 12/2012As an alternative you can also determine the difference of the ORPs of two solutions with one ORP electrode

Página 196

ProLab 3000 ORP voltage93ba75645e03 12/2012Criteria With identical measurement conditions, the following applies:1 In the measured value display, mark

Página 197 - ProLab 3000 Appendix

ORP voltage ProLab 300094ba75645e03 12/20128.2.4 Measurement settingsThe following settings are possible for ORP measurements in the Mea-suring menu:M

Página 198

ProLab 3000 Ion concentration95ba75645e03 12/20129 Ion concentration9.1 General informationTemperature measurement with ion selective measurementsThe

Página 199

Ion concentration ProLab 300096ba75645e03 12/20129.2 Measuring the ion concentration 9.2.1 Preparatory activitiesPerform the following preparatory act

Página 200

ProLab 3000 Ion concentration97ba75645e03 12/20129.2.2 MeasuringTemperature for calibrating and measuringFor precise ion selective measurements, the t

Página 201

Ion concentration ProLab 300098ba75645e03 12/2012Criteria The criteria of the stability control affect the reproducibility of the mea-sured values. I

Página 202

ProLab 3000 Ion concentration99ba75645e03 12/20129.2.3 Measurement settingsAll settings for ion selective measurements are done in the Measuring menu.