ProLab 3000 What to do if...
173
ba75645e03 12/2012
15 What to do if...
15.1 pH and ORP measurement
Error message Err1
Error message Err2
Error message Err4
Error message, Error
Cause Remedy
pH electrode:
Measured value outside the
measuring range
Use a suitable electrode
Air bubble in front of the dia-
phragm
Remove air bubble
Air in the diaphragm Extract air or moisten junc-
tion
Cable broken Replace the electrode
Gel electrolyte dried out Replace the electrode
Cause Remedy
No electrode connected Connect the electrode
Setting time during calibration
too long
Adjust temperature if neces-
sary
Recalibrate
Cause Remedy
Temperature not stable during
calibration.
Adjust temperature if neces-
sary
Recalibrate
Cause Remedy
pH electrode:
The values determined for zero
point and slope of the electrode
are outside the allowed limits.
Recalibrate
Diaphragm contaminated Clean diaphragm