Xylem ProLab 3000 Manual do Utilizador Página 130

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 202
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 129
Ion concentration ProLab 3000
130
ba75645e03 12/2012
13 In the Sample register, enter the value for the sample volume
(Sample volume) if necessary.
14 Use [Dosing] to automatically add to the standard solution the
adjusted volume of the sample (Sample volume).
or
If the burette control is inactive:
Add to the standard solution the adjusted volume of the sample
manually (Sample volume)
15 Start the measurement with [Continue].
The measured value is checked for stability (Stability control).
16 Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Calculation register with the calculated measured value
pops up.
Vista de página 129
1 2 ... 125 126 127 128 129 130 131 132 133 134 135 ... 201 202

Comentários a estes Manuais

Sem comentários