Ion concentration ProLab 3000
122
ba75645e03 12/2012
13 In the Standard register, enter the values for Standard volume
and Standard concentration if necessary.
14 Use [Dosing] to automatically add to the sample the adjusted
volume of the standard solution (Standard volume).
or
If the burette control is inactive:
Add to the sample the adjusted volume of the standard solution
manually (Standard volume)
15 Start the measurement with [Continue].
The measured value is checked for stability (Stability control).
16 Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Calculation register with the calculated measured value
pops up.