ba76014e01 02/2012IQ SENSOR NETFDO®70x IQ (SW)Operating manualOptical IQ SENSOR NET D.O. sensorsFDO® 700 IQFDO® 701 IQFDO® 700 IQFDO® 701 IQ SW
Safety FDO®70x IQ (SW)2 - 2ba76014e01 02/20122.1 Authorized useThe authorized use of the FDO®70x IQ (SW) comprises its use as a D.O. sensor in the IQ
FDO®70x IQ (SW) Safety2 - 3ba76014e01 02/2012Safe operation If safe operation is no longer possible, the sensor must be taken out of operation and sec
Safety FDO®70x IQ (SW)2 - 4ba76014e01 02/2012
FDO®70x IQ (SW) Commissioning3 - 1ba76014e01 02/20123 Commissioning 3.1 Scope of delivery Sensor, consisting of sensor basis and sensor cap Operatin
Commissioning FDO®70x IQ (SW)3 - 2ba76014e01 02/2012Connecting the sensorto the sensorconnection cableFig. 3-1 Connecting the sensor1 Take the protect
FDO®70x IQ (SW) Commissioning3 - 3ba76014e01 02/20123.3 Commissioning / Getting the instrument ready for measuringIdentification in theIQ SENSOR NETBo
Commissioning FDO®70x IQ (SW)3 - 4ba76014e01 02/20123.4 FDO®70x IQ (SW) setting tableSetting Selection/values ExplanationMeasuring mode Concentratio
FDO®70x IQ (SW) Commissioning3 - 5ba76014e01 02/2012Carrying out settings Using s, switch from the measured value display to the main menu of the sett
Commissioning FDO®70x IQ (SW)3 - 6ba76014e01 02/2012
FDO®70x IQ (SW) Measuring / Operation4 - 1ba76014e01 02/20124 Measuring / Operation4.1 MeasuringWARNINGContact with the sample can be dangerous for th
FDO®70x IQ (SW)2ba76014e01 02/2012NoteFor the most recent version of the manual, please visit www.ysi.com.Contact YSI1725 Brannum LaneYellow Springs,
Measuring / Operation FDO®70x IQ (SW)4 - 2ba76014e01 02/2012 With air temperatures under 5 °C we recommend performing the function check and user cal
FDO®70x IQ (SW) Measuring / Operation4 - 3ba76014e01 02/2012Canceling the check As long as the determination of the relative slope has not yet been st
Measuring / Operation FDO®70x IQ (SW)4 - 4ba76014e01 02/20124.2.3 User calibrationPrinciple The user calibration can either be done in water vapor-sat
FDO®70x IQ (SW) Measuring / Operation4 - 5ba76014e01 02/2012Canceling the usercalibrationAs long as the determination of the calibration data has not
Measuring / Operation FDO®70x IQ (SW)4 - 6ba76014e01 02/20124.2.4 Calibration historyCalibration history(available in theIQ SENSOR NET system2020 XT o
FDO®70x IQ (SW) Measuring / Operation4 - 7ba76014e01 02/20124.2.5 Reactivating previous calibration dataThe FDO®70x IQ (SW) enables you to reactivate
Measuring / Operation FDO®70x IQ (SW)4 - 8ba76014e01 02/2012
FDO®70x IQ (SW) Maintenance, cleaning, replacement parts5 - 1ba76014e01 02/20125 Maintenance, cleaning, replacement parts5.1 General maintenance instr
Maintenance, cleaning, replacement parts FDO®70x IQ (SW)5 - 2ba76014e01 02/20125.2 Handling of the sensor capDespite its exterior robustness, the sens
FDO®70x IQ (SW) Maintenance, cleaning, replacement parts5 - 3ba76014e01 02/2012Mounting the sensorcap5.4 Cleaning the sensor5.4.1 Exterior cleaningDir
FDO®70x IQ (SW) Contents0 - 1ba76014e01 02/2012FDO®70x IQ (SW) - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Maintenance, cleaning, replacement parts FDO®70x IQ (SW)5 - 4ba76014e01 02/20125.4.2 Interior cleaning of sensor cap and sensor headIf moisture or dir
FDO®70x IQ (SW) Maintenance, cleaning, replacement parts5 - 5ba76014e01 02/20125.5 Replacement parts and accessoriesComponents forcompressed air clean
Maintenance, cleaning, replacement parts FDO®70x IQ (SW)5 - 6ba76014e01 02/2012
FDO®70x IQ (SW) What to do if...6 - 1ba76014e01 02/20126 What to do if...Sensor does not appearin the measured valuedisplay and list ofsensorsMeasured
What to do if... FDO®70x IQ (SW)6 - 2ba76014e01 02/2012Measured value invalid( "----" displayed)Incorrect temperaturedisplayCause Remedy– Us
FDO®70x IQ (SW) Technical data7 - 1ba76014e01 02/20127 Technical data7.1 General measurement characteristicsNoteThe measurement characteristics are pr
Technical data FDO®70x IQ (SW)7 - 2ba76014e01 02/20127.3 General dataDimensionsWeight (without sensorconnection cable)Connection technique Connection
FDO®70x IQ (SW) Technical data7 - 3ba76014e01 02/2012Automatic sensormonitoring(SensCheck function)Monitoring of the membrane functionInstrument safet
Technical data FDO®70x IQ (SW)7 - 4ba76014e01 02/20127.5 Technical data of the SC-FDO®700Measuring ranges andresolutionsD. O. partial pressure 0 ... 4
FDO®70x IQ (SW) Technical data7 - 5ba76014e01 02/20127.6 Technical data of the SC-FDO®701Measuring ranges andresolutionsD. O. partial pressure 0 ... 4
Contents FDO®70x IQ (SW)0 - 2ba76014e01 02/20127 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-17.1 General measure
Technical data FDO®70x IQ (SW)7 - 6ba76014e01 02/2012
FDO®70x IQ (SW) Contact Information8 - 1ba76014e01 02/20128 Contact Information8.1 Ordering & Technical SupportWhen placing an order please have t
Contact Information FDO®70x IQ (SW)8 - 2ba76014e01 02/2012
FDO®70x IQ (SW) Indexes9 - 1ba76014e01 02/20129 Indexes9.1 Explanation of the messagesThis chapter contains a list of all the message codes and relate
Indexes FDO®70x IQ (SW)9 - 2ba76014e01 02/2012EC833x Sensor could not be calibrated,sensor blocked for measurementCause: instable signal* Check temper
FDO®70x IQ (SW) Indexes9 - 3ba76014e01 02/20129.1.2 Info messagesMessage code Message textIC133x Sensor has been successfully calibrated* For calibrat
Indexes FDO®70x IQ (SW)9 - 4ba76014e01 02/20129.2 Status infoThe status info is a coded piece of information on the current status of a sensor. Each s
1725 Brannum LaneYellow Springs, Ohio 45387 USA+1 937-767-7241800-765-4974 (US)FAX (937) 767-1058Email: [email protected]: www.ysi.com
FDO®70x IQ (SW) Overview1 - 1ba76014e01 02/20121Overview1.1 How to use this component operating manualStructure of theIQ SENSOR NEToperating manualFig
Overview FDO®70x IQ (SW)1 - 2ba76014e01 02/20121.2 ModelsThe FDO®70x IQ (SW) consists of a sensor basis equipped with differ-ent sensor cap types, dep
FDO®70x IQ (SW) Overview1 - 3ba76014e01 02/20121.3 Recommended fields of applicationSensor model Recommended field of application.FDO®700 IQFDO®701 IQ
Overview FDO®70x IQ (SW)1 - 4ba76014e01 02/20121.4 StructureStructureFig. 1-2 Structure of the D. O. sensor (example: FDO® 70x IQ)Sensor cap withmemor
FDO®70x IQ (SW) Safety2 - 1ba76014e01 02/20122 SafetyThis component operating manual contains special instructions that must be followed during the op
Comentários a estes Manuais