Xylem CellOx 325 Dissolved oxygen sensor Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Ferramentas Xylem CellOx 325 Dissolved oxygen sensor. Xylem CellOx 325 Dissolved oxygen sensor User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 16
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 0
CellOx 325
17
Operating manual
Dissolved oxygen sensor
Vista de página 0
1 2 3 4 5 6 ... 15 16

Resumo do Conteúdo

Página 1 - CellOx 325

CellOx 32517Operating manualDissolved oxygen sensor

Página 2 - Printed in Germany

Maintenance, cleaning, replacement CellOx 325265.4 Cleaning the electrodesCautionBefore starting to work with the sensor, please note the GENERAL MAIN

Página 3 - CellOx 325 - Contents

CellOx 325 Maintenance, cleaning, replacement27Thoroughly rinse the sensor head with deionized water.Water the counter electrode in deionized water fo

Página 4 - 1 Overview

Maintenance, cleaning, replacement CellOx 325285.5 Checking the sensor for freedom from zero-currentThe sensor is zero-current free. Checking the sens

Página 5 - 3 Commissioning

CellOx 325 What to do if...295.6 DisposalCautionThe ELY/G electrolyte solution irritates eyes and skin. When dealing with the ELY/G electrolyte soluti

Página 6 - 4 Measuring / Operation

Technical data CellOx 325307 Technical dataGeneral features Measuring principle Membrane covered galvanic sensorTemperature compensation IMT compensat

Página 7 - 5.2 Exterior cleaning

CellOx 325 Technical data31Pin assignmentMeasurementconditionsMeasuring ranges at 20 °C 0 ... 50 mg/l O20 ... 600 % O2 saturation0 ... 1250 mbar O2 pa

Página 8 - RL/G H O

Wear parts and accessories CellOx 325328 Wear parts and accessoriesNoteFor further accessories, refer to the WTW catalog or the Internet.Wear parts an

Página 9

CellOx 32518Copyright © Weilheim 2002, WTW GmbH & Co. KGReprinting - even as excerpts - is only allowed with the explicit written autho-rization o

Página 10 - 5.4 Cleaning the electrodes

CellOx 325 Contents19CellOx 325 - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .202 Safety . . . . .

Página 11

Overview CellOx 325201 OverviewConstructionRecommendedfields ofapplicationl On site measurements in rivers, lakes and wastewaterl Applications in wate

Página 12

CellOx 325 Safety212SafetyThis operating manual contains special instructions that must be followed during the operation of the Dissolved Oxygen (D.O.

Página 13 - 6 What to do if

Measuring / Operation CellOx 325224 Measuring / Operation4.1 CalibrationNoteFor calibration, please refer to the operating manual of the measuring ins

Página 14 - 7 Technical data

CellOx 325 Maintenance, cleaning, replacement235 Maintenance, cleaning, replacement5.1 General maintenance instructionsFor your safety When dealing wi

Página 15 - CellOx 325 Technical data

Maintenance, cleaning, replacement CellOx 325245.3 Changing the electrolyte and membrane headCautionBefore starting to work with the sensor, please no

Página 16 - 8 Wear parts and accessories

CellOx 325 Maintenance, cleaning, replacement25Carefully shake off the drops of water.Fill a new membrane cap with ELY/G electrolyte solution. Remove

Comentários a estes Manuais

Sem comentários