Safety Cond 311010ba75792e02 11/20082.1 Authorized useAuthorized use of the meter consists exclusively of the measurement of conductivity, temperature
Cond 3110 Safety11ba75792e02 11/2008Safe operation If safe operation is no longer possible, the meter must be taken out of service and secured against
Safety Cond 311012ba75792e02 11/2008
Cond 3110 Commissioning13ba75792e02 11/20083 Commissioning3.1 Scope of deliveryz Conductivity meter Cond 3110 z 4 batteries 1.5 V Mignon type AAz Oper
Commissioning Cond 311014ba75792e02 11/2008CautionMake sure that the poles of the batteries are the right way round.The ± signs on the batteries must
Cond 3110 Operation15ba75792e02 11/20084 Operation4.1 General operating principlesThis section contains basic information on the operation of the Cond
Operation Cond 311016ba75792e02 11/20084.2 MeasuringPreparatory activities Perform the following preparatory activities when you want to measure:Stabi
Cond 3110 Operation17ba75792e02 11/20084.2.1 Measuring the conductivity4.2.2 Measuring the salinity 1 Perform the preparatory activities according to
Operation Cond 311018ba75792e02 11/2008212323C°SalnLFTPARARng
Cond 3110 Operation19ba75792e02 11/20084.3 Determining/setting up the cell constant [C]Why determine/set upthe cell constant?Due to aging, the cell co
Cond 3110ba75792e02 11/2008Accuracy whengoing to pressThe use of advanced technology and the high quality standard of our instruments are the result o
Operation Cond 311020ba75792e02 11/2008NoteIf the error message E3 appears, refer to chapter 6 WHAT TO DO IF...Stability control During calibration, t
Cond 3110 Operation21ba75792e02 11/2008Calibration evaluation After the calibration, the meter automatically evaluates the current status. The evaluat
Operation Cond 311022ba75792e02 11/20084.3.3 Setting the cell constant manuallyNoteThe cell constant to be set must either be taken from the operating
Cond 3110 Operation23ba75792e02 11/20084 Set the cell constant to be used with <▲><▼>, e.g. 0.837 cm-1.5 Confirm the selection with <EN
Operation Cond 311024ba75792e02 11/2008Selecting the cellconstant0.475 cm-11 Press the <CAL> key repeatedly until USE CELL is displayed.2 Confir
Cond 3110 Operation25ba75792e02 11/20084.4 Temperature compensation TCThe calculation of the temperature compensation is based on the preset reference
Operation Cond 311026ba75792e02 11/20084.5 SettingsYou can adapt the meter to your individual requirements. The settings are done in the following men
Cond 3110 Operation27ba75792e02 11/20084.5.2 Measurement settingsThese settings apply to the determination of the cell constant and measurement (the d
Operation Cond 311028ba75792e02 11/2008Cleaning interval (Int.C) 4 Using <▲><▼>, toggle between °C and °F.5 Confirm with <ENTER>.Int
Cond 3110 Operation29ba75792e02 11/20084.6 Reset4.6.1 Resetting the cell constantThis function serves to erase the last determined cell constant. Subs
Cond 3110 Contents3ba75792e02 11/2008Cond 3110 - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.1 K
Operation Cond 311030ba75792e02 11/20084.6.2 Resetting all meter settingsThis function resets all meter settings to the default condition. The relevan
Cond 3110 Maintenance, cleaning, disposal31ba75792e02 11/20085 Maintenance, cleaning, disposal5.1 MaintenanceThe only maintenance activity required is
Maintenance, cleaning, disposal Cond 311032ba75792e02 11/2008CautionMake sure that the poles of the batteries are the right way round.The ± signs on t
Cond 3110 What to do if...33ba75792e02 11/20086 What to do if...Error messageOFL, UFLError message,E3Sensor symbol flashesDisplay,LoBatMeter does not
What to do if... Cond 311034ba75792e02 11/2008
Cond 3110 Technical data35ba75792e02 11/20087 Technical data7.1 General data7.2 Measuring ranges, resolution, accuracyDimensions approx. 240 x 190 x 8
Technical data Cond 311036ba75792e02 11/2008NoteThe accuracy values specified here apply exclusively to the meter. The accuracy of the measuring cell
Cond 3110 Lists37ba75792e02 11/20088ListsThis chapter provides additional information and orientation aids.Specialist terms The glossary briefly expla
Lists Cond 311038ba75792e02 11/2008Reset Restoring the original condition of all settings of a measuring system.Resistance Short name for the specific
Cond 3110 ListsIndex39ba75792e02 11/2008AAuthorized use . . . . . . . . . . . . . . . . . . . . . . 10AutoRead . . . . . . . . . . . . . . . . . . . .
Contents Cond 31104ba75792e02 11/20086 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 337 Technical data . . . . .
Lists Cond 311040ba75792e02 11/2008
Wissenschaftlich-Technische Werkstätten GmbHDr.-Karl-Slevogt-Straße 1D-82362 WeilheimGermanyTel: +49 (0) 881 183-0+49 (0) 881 183-100Fax: +49 (0) 881
Cond 3110 Overview5ba75792e02 11/20081OverviewThe Cond 3110 compact precision conductivity meter enables you to perform conductivity measurements quic
Overview Cond 31106ba75792e02 11/20081.1 KeypadIn this operating manual, keys are indicated by brackets <..> . The key symbol (e.g. <ENTER>
Cond 3110 Overview7ba75792e02 11/20081.2 DisplayStatus displayindicatorsTime1/cm/K%88888888°FC°SalS/cmmLoBatTref25Tref20ARngCalnLF ARTP-µMeasured para
Overview Cond 31108ba75792e02 11/20081.3 Socket fieldCautionOnly connect sensors to the meter that cannot return any voltages or currents that are not
Cond 3110 Safety9ba75792e02 11/20082 SafetyThis operating manual contains basic instructions that you must follow during the commissioning, operation
Comentários a estes Manuais