Resumo do Conteúdo

Página 1 - User's Guide

Dissolved OxygenDO600IIRECALLUser's Guide ExStik® DO600 Dissolved Oxygen Meter

Página 2 - Meter Description

DO600 - Version 5.0 8/05 10 Operational Matrix for the Model DO600 Function / Resulting Action Power Status Mode Setting Required Button Press Sequen

Página 3 - Operation

DO600 - Version 5.0 8/05 11  Support line (781) 890-7440 Technical support: Extension 200; E-mail: [email protected] Repair & Returns: Extensio

Página 4 - Temperature Units (ºF / ºC)

DO600 - Version 5.0 8/05 2 Dissolved OxygenDO600IIRECALL123456789MembraneCathodeElectrode Bottom ViewIntroduction Congratulations on your purchase of

Página 5

DO600 - Version 5.0 8/05 3 Operation® Powering the ExStik® The ExStik® uses four (4) CR2032 Lithium Ion Batteries. Press the ON/OFF button to turn th

Página 6 - Maintenance

DO600 - Version 5.0 8/05 4 Calibration 1. Calibration should be performed on a daily basis. 2. Turn the meter ON 3. Press and hold the MODE/HOLD

Página 7 - Counter-clockwise to loosen

DO600 - Version 5.0 8/05 5 Altitude Compensation 1. With the unit ON, momentarily press the CAL/RECALL button twice in quick succession (‘SAL’ is sh

Página 8 - Specifications

DO600 - Version 5.0 8/05 6 Measurement and Display Considerations • The DO600 uses a polarographic sensor, which consumes oxygen at the sensing surf

Página 9 - Appendices

DO600 - Version 5.0 8/05 7 BONDED CAPClockwise to tightenCounter-clockwise to loosenDO Bonded Membrane Cap replacement Important Note : Do not touch

Página 10 - DO600 - Version 5.0 8/05

DO600 - Version 5.0 8/05 8 Specifications Display 2000 count, Dual function 3 ½ digit LCD with Bargraph, Display size: 24 mm x 20 mm Sensor Polar

Página 11 - Warranty

DO600 - Version 5.0 8/05 9 Appendices Troubleshooting Guide Symptom Possible Causes Action Unit will not power on -Batteries not in place -Dead bat