Global Water
MODBUS MANUAL
For the
CL500
Online Residual
Chlorine Monitor
Global Water
11390 Amalgam Way
Gold River CA 95670
Phone: 800-876-1172
Fax: 847-672-9988
Website: www.globalw.com
Revised (3/09)
Rev. 1.3
Vista de página 0
1 2 3 4 5 6 ... 13 14

Resumo do Conteúdo

Página 1 - Global Water

Global Water MODBUS MANUAL For the CL500 Online Residual Chlorine Monitor Global Water 11390 Amalgam Way Gold

Página 2

4.3 Holding Registers These 16-bit values are readable and changeable from the master. The data is stored and transmitted with the MSB first and then

Página 3

Address Type Register Min Default Max Function 2 -- Odd 40013 Int RS-485 stop bits 0 0 1 0 – One 1 -- Two 40014 Int Instrument address 1 1 255

Página 4

4.4 Input Registers These 16-bit values are readable by the master. The data is stored with the MSB first and then the LSB. 4.4.1 Valid Command(s) Co

Página 5

0x00000 Normal 0x00001 Break in the 4-20 mA current loop (MA) 0x00002 Calibration error (ZCAL) 0x00004 No intake water (WATR) 0x00008 Intake water

Página 6 - 3.0 Operation

CL500 Modbus (3/09) Rev. 1.3 12Note: Error messages shown in Prentiss and in (Italics) 4.5 Exception Responses Implemented Code Name Meaning 00 --

Página 8 - 4.1 Coils

1.0 Overview The Global Water CL500 uses a communication protocol called Modbus. A company called Modicon, for use with their programmable contro

Página 9 - 4.2 Input Status

2.0 Electrical Connections All of the electrical connections to the instrument are made at the termination area, which is located on the portion

Página 10 - 4.3 Holding Registers

2.1 RS-485 Connection The RS-485 half-duplex (2-wire) digital interface operates with differential levels that are not susceptible to electrical inte

Página 11

3.0 Operation 3.1 Configuring the RS–485 Port The CL500 is equipped with an RS-485 port which operates in one of three ways (see the operator’s ma

Página 12 - 4.4 Input Registers

3.1 Configuring the RS–485 Port (continued) For Modbus some applications, especially involving the ASCII operation mode, it may be necessary to set

Página 13

4.0 The Modbus RS-485 Output & Commands Implemented The default communication parameters are 8 bits, no parity and 1 stop bit. Please note tha

Página 14

4.2 Input Status These single-bit values are readable from the master. The data will be returned with the lowest addressed input status in the LSB of