ba75825e01 08/2009LR 925/01Operating manualUltrapure water conductivity measuring cell
LR 925/01ba75825e01 08/2009NoteThe latest version of the present operating manual can be found on the Internet under www.WTW.com.Copyright © Weilheim
LR 925/01 Contents11ba75825e01 08/2009LR 925/01 - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Overview LR 925/0112ba75825e01 08/20091Overview1.1 Structure and functionStructureAutomatic sensorrecognitionThe sensor electronics with the stored se
LR 925/01 Cleaning13ba75825e01 08/20092 CleaningCAUTIONTo clean the sensor, disconnect it from the instrument.Outside cleaning We recommend to clean t
Technical data LR 925/0114ba75825e01 08/20094 Technical data4.1 General dataGeneral features Measuring principle 2-electrodes measurementCell constant
LR 925/01 Technical data15ba75825e01 08/20094.2 Measuring ranges and resolutionMeasurementconditionsConductivity measuring range0.01 μS/cm ... 200 μS/
Technical data LR 925/0116ba75825e01 08/20094.3 Accuracy of the IDS measuring techniqueMeasured parameter Accuracy (± 1 digit)ӕ, ρ ± 0.5 % of measured