6-Series Multiparameter Water Quality Sondes User Manual 6-Series: 6600 V2 6600EDS V2 6920 V2 682
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-4 Figure 4 Sonde to 650 Display/Logger You will need
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-94 You may choose the entire file. Use the Space Bar to a
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-95 2.9.4 STATUS Select 4-Status from the Sonde Main menu
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-96 2.9.5 SYSTEM Select 5-System from the Sonde Main menu
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-97 The default is 9600, but you may change it to match you
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-98 The asterisks (*) that follow the numbers or letters i
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-99 Oxygen in % air saturation, Dissolved Oxygen in mg/L, p
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-100 PAR1 Output from special photosynthetically active ra
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-101 As noted above, the ISE3 PAR1 selection is used in a s
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-102 Select 1-Cal constants to display the calibration con
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-103 Turb+ A2 1000 0.6 to 1.5 Range is ratio of (M2
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-5 2.3 PREPARING THE SONDE FOR USE To prepare the sonde
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-104 Users should select the option for their
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-105 6-( ) Multi SDI12. Modifies the SDI12 protocol as fo
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-106 NOTE: The number of items on this menu depends great
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-107 calculation of depth or level. This item will only app
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-108 activated at the interval assigned in the Unattended s
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-109 Sondes with no optical (turbidity, chlorophyll, or rho
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-110 Choosing 4-Threshold will display the following:
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-111 reading in air is near zero, and quite likely, is a ve
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-112 When caring for your sonde, remember that the sonde is
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-113 SONDE PROBE PORTS Whenever you install, remove or r
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-6 DO MEMBRANE INSTALLATION WITH THE PROBE NOT INSTALLED
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-114 To resurface the probe using the fine sanding disk, f
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-115 YSI recommends the optical DO membrane assembly be re
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-116 Membrane Installation Instructions Use the schematic
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-117 Store the probe with new membrane in either water o
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-118 2. Rinse the probe in clean tap water, wipe with a co
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-119 regeneration cycle has been successful. The felt filt
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-120 2.10.3 SONDE STORAGE Proper storage of the your s
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-121 plug. Leave the conductivity/temperature and the diss
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-122 RAPID PULSE DISSOLVED OXYGEN Rapid Pulse dissolved
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-123 ROX OPTICAL DISSOLVED OXYGEN When the 6150 sensor i
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-7 DO MEMBRANE INSTALLATION WITH THE PROBE INSTALLED IN THE
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-124 AMMONIUM, NITRATE AND CHLORIDE The active element in
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-1 SECTION 3 650 MDS DISPLAY/LOGGER 3.1 INTRODUCTION The YSI 650 M
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-2 3.2.1 UNPACKING Remove the instrument from the shipping box, being
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-3 used to log data in Discrete or Unattended sampling applications.
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-4 Figure 2. 650 interfaced to a 6-series sonde with user-supplied GPS
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-5 Figure 4. Charging of 650 rechargeable battery pack. The setup o
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-6 3.2.4 650 FEATURES The key physical features of the 650 display a
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-7 Note that the back of the case contains the battery lid that is a
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-8 The circuit board in this instrument contains a manganese dioxide li
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-9 Make certain that the gasket is properly installed on the rechar
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-8 your fingers. Once the O-ring is in position, squeeze i
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-10 To use the device with an American or Japanese vehicle, slide th
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-11 Note that the screen is divided into two sections – the “Main Displ
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-12 3.2.8 USING THE YSI 650 KEYPAD The
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-13 1 -- 1 2 – ABC2abc2 3 – DEF3def3 4 – GHI4ghi4 5 – JKL5jkl5 6 – MNO6
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-14 Use the arrow keys to highlight the Date and Time selection and pre
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-15 As shown above, enter the designation “
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-16 “click”. After connecting the sonde to the 650, press the “on” but
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-17 Section 3.11 of this manual. If symptoms persist after these acti
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-18 3.2.11 UPGRADING ECOWATCH FOR WINDOWS SOFTWARE
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-19 YSI suggests that the user become familiar with th
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-9 REMOVING THE PORT PLUGS Using the long extended end of
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-20 3.3.7 ID Activate this item and Enter choice of instrument identi
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-21 3.4 SONDE MENU INTERFACE 3.4.1 INTRODUCTION The Sonde menu se
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-22 In fact, the only significant difference between the two interfac
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-23 3.4.2 SONDE MENU EXAMPLES To become familiar with the 650 Sonde
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-24 Use the arrow keys to highlight the Dissolved Oxyge
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-25 “Calibrate” to “Continue”. Make sure that “Continue” is highlighte
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-26 3.4.3.4 Calibrating sonde sensors Use the 650 Sonde menu selection
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-27 The rules are particularly important with regard to the method of d
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-28 The user should also
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-29 Decide if you want to log data continuously, and, if so, determ
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-10 600XL & 600XLM SONDE BULKHEAD 3 Port Sonde: 1 Rapi
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-30 The user places the sonde in the water and watches the readings on
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-31 The user attaches the 6920 sonde to a 650. After turning on the 6
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-32 At the end of the 90 minute profile determination, the
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-33 3.5.3 LOGGING DATA TO 650 MEMORY 3.5.3.1 INTRODUCTION Even if
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-34 Press Enter at the highlighted Interval selection and use the arrow
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-35 Using the arrow keys, the Filename window is highlighted, the user
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-36 3.5.3.3 LOGGING TO 650 MEMORY WITH A SITE LIST In the example abov
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-37 Note that Site lists containing Single Site Des
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-38 Note also that if you have purchased a 650 without the b
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-39 Then press Enter. When the logging prompt is activated,
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-11 6600EDS V2-2 SONDE BULKHEAD 5 Port Sonde: 1 Rapid Pulse
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-40 These files will remain “open” in 650 memory so that yo
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-41 Then highlight the Edit site list selection and press
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-42 Then move to the next location (EAST), activate the 650 logging fun
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-43 Note that the data from two sites is identified by the Site Numbe
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-44 The two inserted entries can be moved either up or down
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-45 3.5.4 650 LOGGING – “CANS” AND “CAN’TS” The use of the 650 to fa
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-46 3.5.4.2 SUMMARY OF 650 LOGGING LIMITATIONS WITH A 650/YSI SONDE SY
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-47 To proceed with the details of 650 File management, turn the instru
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-48 For both files, the file name, site name, file size and time of th
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-49 3.6.3 UPLOAD TO PC This frequently used command is used to trans
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-12 6820V2-1 & 6920V2-1 SONDE BULKHEADS 7 Por
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-50 To transfer data from the 650 to your PC, highlight the Upload to P
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-51 Use the arrow keys to scroll vertically to view more data points
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-52 3.7 UPLOADING DATA FROM SONDES 3.7.1 INTRODUCTION A primary f
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-53 3.8 USING GPS WITH THE 650 3.8.1 SETTING UP THE GPS-650 INTERF
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-54 The GPS readings must be logged to 650 memory from the 650 Run d
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-55 3.9 USING THE 650 BAROMETER 3.9.1 PRINCIPLES OF OPERATION The
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-56 3.9.4 LOGGING BAROMETER READINGS To log barometric
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-57 3.10 UPGRADING 650 SOFTWARE The 650 software can easily be upgra
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-58 3.11 TROUBLESHOOTING The following sections describes problems w
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-59 3.12 FERRITE BEAD INSTALLATION WARNING: If you are using your 65
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-13 600LS BULKHEAD Fi
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-60 3.13 650 SAFETY CONSIDERATIONS IMPORTANT SAFETY INSTRUCTIONS! S
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-61 Danger Notifications – Misuse creates a STRONG possibility of death
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-62 616 Cigarette Lighter Charger Safety Information 1. This section
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-63 CAUTION: If water has contacted the rechargeable battery pack, do n
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-64 Barometer: Range: 500 to 800 mm Hg Resolution: 0.1 mm Hg Accuracy:
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-65
650 MDS Section 3 YSI Environmental Monitoring Systems Operations Manual 3-66
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-1 SECTION 4 ECOWATCH FOR WINDOWS 4.1 INTRODUCTIO
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-2 6 - Page The Data Display Window can be divided in
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-3 EcoWatch starts without an open data file (or .DAT fi
SAFETY NOTES TECHNICAL SUPPORT AND WARRANTY INFORMATION Contact information for technical support and warranty information on YSI’s Environmental
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-14 600 OMS V2-1 BULKHEAD
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-4 Click To: Open an existing data file (.DAT). Eco
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-5 4.1.5 TYPICAL APPLICATION Suppose you want to mea
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-6 5. Print Your Graph To print the graph, open the F
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-7 4.2 DATA ACQUISTION AND ANALYSIS 4.2.1 CONNECT A Y
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-8 A sonde or 650 display/logger must be connected to o
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-9 To transfer data from the 650 to your PC, highlight
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-10 Unzoom Use this button to view the study from
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-11 4.2.5 GRAPH DATA IN REAL TIME Choosing the New co
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-12 4.3 ECOWATCH MENU 4.3.1 FILE The File menu off
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-13 CLOSE Use this command to close all windows contai
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-15 INSTALLING THE TURBIDITY, CHLOROPHYLL, RHODAMINE WT
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-14 COPY TO CLIPBOARD If a graph or table is active,
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-15 SAVE DATA DISPLAY When opening a data file, you wil
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-16 EXPORT Use this command to send the current dat
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-17 individual data entries, quotation marks surround an
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-18 PRINT Use this command to print a document. If a gr
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-19 PRINT SETUP Use this command to select a printer
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-20 Orientation Choose Portrait or Landscape. Paper Siz
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-21 4.3.2 EDIT The edit menu appears whenever a data
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-22 INSERT PARAMETER Adds an additional parameter to th
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-23 NUM The Num Lock key is latched down. SCRL T
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-16 INSTALLING THE ISE PROBES Fi
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-24 Shortcuts Toolbar: STUDY INFO This command wi
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-25 ZOOM IN Enlarges graph or data table (whichever i
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-26 Port Setup This is where you can set port param
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-27 If a modem is selected for a particular COM port,
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-28 Dialing If there is no entry for Phone Number, th
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-29 Default Sonde Port This refers to the COM port yo
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-30 A common example is using a sonde in Run Mode. The
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-31 4.3.5 REAL-TIME The Real-time menu offers the fo
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-32 4.3.6 GRAPH ZOOM WINDOW Enlarges a section of
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-33 MANUAL SCALE Takes you to the Graph Y-Axis dialog
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-17 2.3.3 STEP 3 - POWER Some type of external power su
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-34 Add/Remove There are two lists, Selected Parameter
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-35 Attributes This command takes you to the Change
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-36 Calibrate TSS TSS is an abbreviation for Total
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-37 Once you have at least two correlation points, you h
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-38 Color Type or select one of the 16 predefined col
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-39 Note that there are two color buttons, one for the t
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-40 1 - Open a data file and observe the presentation
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-41 4.3.9 WINDOW The Window menu offers the follow
EcoWatch for Windows Section 4 YSI Incorporated Environmental Monitoring Systems Manual 4-42 4.3.10 HELP The Help menu offers the following co
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-1 SECTION 5 PRINCIPLES OF OPERATION 5.1 CONDUCTIVI
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-18 INSTALLING BATTERIES The 600XLM, 6600V2-2, 6600EDS V
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-2 4 - Perform sensor calibration at a temperature a
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-3 generally speaking, TDS cannot be calculated accur
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-4 MEASUREMENT AND CALIBRATION PRECAUTIONS 1 - Ins
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-5 CALIBRATION AND EFFECT OF TEMPERATURE The softw
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-6 4 - During pH calibration, allow the sensors time
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-7 MEASUREMENT AND CALIBRATION PRECAUTIONS (1) Be ce
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-8 Teflon membrane. The current associated with this
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-9 protocol according to the instructions. The calib
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-10 (4) For short term storage (2 weeks or less), k
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-11 EFFECT OF TEMPERATURE
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-19 INSTALLING BATTERIES INTO THE YSI 6600V2-2, 6600EDS V2-
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-12 MEASUREMENT AND CALIBRATION PRECAUTIONS (1) If
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-13 change in voltage. Our experience with ISEs indi
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-14 use the less time-consuming 2 point and 1-point r
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-15 = 0.6). Even if the ionic content were due to so
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-16 (2) Ion selective electrodes may not stabilize as
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-17 instrument can be frequently calibrated. However
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-18 sonde software to provide readings in nephelometr
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-19 turbidity standards supplied by YSI, the value of
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-20 example, the label of the YSI 6073 turbidity stan
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-21 For most unattended sampling applications, select
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-20 2. Press two of the outermost retaining tabs down into
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-22 chlorophyll system are designed to complement the
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-23 6025 chlorophyll probe is equipped with a mechani
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-24 PREPARATION OF DYE SOLUTIONS FOR CHECKING SENSOR
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-25 Table. Approximate algal chlorophyll equivalent o
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-26 wiper that makes it ideal for unattended applicat
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-27 standard that provides the best measure of accura
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-28 content of an environmental sample may appear to
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-29 Field optical measurements are particularly su
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-30 RHODAMINE WT STANDARD SOLUTION – PREPARATION AND
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-31 experiments indicate that a suspension of phytopl
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-21 cannot be reset and the battery lid must be replaced be
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-32 described above. The YSI 6131 sensor is designed
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-33 CALIBRATION METHODS For calibration of the YSI B
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-34 2. Accurately weigh 0.500 g of the 20 % Rhodamin
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-35 EFFECT OF CHLOROPHYLL ON BGA-PC READINGS While t
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-36 capability of mechanical cleaning is likely to be
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-37 5.17 PHYCOERYTHRIN-CONTAINING BLUE-GREEN ALGAE
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-38 Measurement of PE-containing BGA In Vivo One ke
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-39 CALIBRATION METHODS For calibration of the YSI B
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-40 2. Accurately weigh 0.500 g of the 20 % Rhodamin
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-41 EFFECT OF CHLOROPHYLL ON BGA-PE READINGS While
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-22 Check the O-rings and sealing surfaces for any contamin
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-42 mechanical wiper that makes it ideal for unattend
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-43 n 5.18 FLOW Flow is a calculated value. Whene
Principles of Operation Section 5 YSI Incorporated Environmental Monitoring Systems Manual 5-44 vegetation in the channel also affects the value
Troubleshooting Section 6 YSI Incorporated Environmental Monitoring Systems Operations Manual 6-1 SECTION 6 TROUBLESHOOTING This section co
Troubleshooting Section 6 YSI Incorporated Environmental Monitoring Systems Operations Manual 6-2 6.2 SONDE COMMUNICATION PROBLEMS SYMPTOM POSSI
Troubleshooting Section 6 YSI Incorporated Environmental Monitoring Systems Operations Manual 6-3 6.3 SENSOR PERFORMANCE PROBLEMS SYMPTOMS POSSI
Troubleshooting Section 6 YSI Incorporated Environmental Monitoring Systems Operations Manual 6-4 Water in probe connector Dry connector; reinst
Communications Section 7 YSI Incorporated Environmental Monitoring Systems Operations Manual 7-1 SECTION 7 COMMUNICATION This section describes
Communications Section 7 YSI Incorporated Environmental Monitoring Systems Operations Manual 7-2 7.3 RS-232 INTERFACE The sonde has an auto-baud
Communications Section 7 YSI Incorporated Environmental Monitoring Systems Operations Manual 7-3 ttt - Maximum time in seconds the Sonde will ta
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-23 CAUTION: The 6067B cable is for laboratory use only --
Communications Section 7 YSI Incorporated Environmental Monitoring Systems Operations Manual 7-4 Manufacturer: YSIIWQSG Model: EMS60
Upgrading Sonde Firmware Section 8 YSI Incorporated Environmental Monitoring Systems Operations Manual 8-1 SECTIO
Upgrading Sonde Firmware Section 8 YSI Incorporated Environmental Monitoring Systems Operations Manual 8-2
Warranty and Service Information Section 9 YSI Incorporated Environmental Monitoring Systems Operations Manual 9-1
Warranty and Service Information Section 9 YSI Incorporated Environmental Monitoring Systems Operations Manual 9-2
Warranty and Service Information Section 9 YSI Incorporated Environmental Monitoring Systems Operations Manual 9-3
Warranty and Service Information Section 9 YSI Incorporated Environmental Monitoring Systems Operations Manual 9-4
Warranty and Service Information Section 9 YSI Incorporated Environmental Monitoring Systems Operations Manual 9-5
Health and Safety Appendix A YSI Incorporated Environmental Monitoring Systems Operations Manual A-1 APPENDIX A HEALTH AND SAFETY NOTE: For a
Health and Safety Appendix A YSI Incorporated Environmental Monitoring Systems Operations Manual A-2 pH 4 INGREDIENTS: θ Potassium Hyd
TABLE OF CONTENTS SECTION 1 INTRODUCTION 1.1 ABOUT YSI 1-1 1.2 HOW TO USE THIS MANUAL 1-1 1.3 UNPAC
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-24 2.4 ECOWATCH FOR WINDOWS -GETTING STARTED This sectio
Health and Safety Appendix A YSI Incorporated Environmental Monitoring Systems Operations Manual A-3 INGREDIENTS: θ Potassium Chloride θ
Health and Safety Appendix A YSI Incorporated Environmental Monitoring Systems Operations Manual A-4 INGREDIENTS: θ Ammonium Chloride θ
Health and Safety Appendix A YSI Incorporated Environmental Monitoring Systems Operations Manual A-5 INGREDIENTS θ Potassium Nitrate θ
Health and Safety Appendix A YSI Incorporated Environmental Monitoring Systems Operations Manual A-6 INGREDIENTS θ Styrene divinylbenzene c
Required Notice Appendix B YSI Incorporated Environmental Monitoring Systems Operations Manual B-1 APPENDIX B REQUIRED NOTICE The Federal Com
Required Notice Appendix B YSI Incorporated Environmental Monitoring Systems Operations Manual B-2
Accessories and Calibration Standards Appendix C YSI Incorporated Environmental Monitoring Systems Operations Manual C-1 APPENDIX C ACCESSORIES
Accessories and Calibration Standards Appendix C YSI Incorporated Environmental Monitoring Systems Operations Manual C-2 θ 6628 Wiper Kit for 66
Accessories and Calibration Standards Appendix C YSI Incorporated Environmental Monitoring Systems Operations Manual C-3 CONDUCTIVITY REAGENTS
Accessories and Calibration Standards Appendix C YSI Incorporated Environmental Monitoring Systems Operations Manual C-4 θ 650-04 – High memory
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-25 2.5 SONDE SOFTWARE SETUP There are two sets of softw
Solubility and Pressure/Altitude Tables Appendix D YSI Incorporated Environmental Monitoring Systems Operations Manual D-1 APPENDIX D SOLUBILITY
Solubility and Pressure/Altitude Tables Appendix D YSI Incorporated Environmental Monitoring Systems Operations Manual D-2 τ
Solubility and Pressure/Altitude Tables Appendix D YSI Incorporated Environmental Monitoring Systems Operations Manual D-3 Table 2: Calibration V
Solubility and Pressure/Altitude Tables Appendix D YSI Incorporated Environmental Monitoring Systems Operations Manual D-4 Table 3. Conversion Fa
Turbidity Measurements Appendix E YSI Incorporated Environmental Monitoring Systems Operations Manual E-1 APPENDIX E TURBIDITY MEASUREMENTS Th
Turbidity Measurements Appendix E YSI Incorporated Environmental Monitoring Systems Operations Manual E-2 not recommend that you use tap water as
Turbidity Measurements Appendix E YSI Incorporated Environmental Monitoring Systems Operations Manual E-3 For most applications involving both s
Turbidity Measurements Appendix E YSI Incorporated Environmental Monitoring Systems Operations Manual E-4 How do I manually activate the turbidit
Turbidity Measurements Appendix E YSI Incorporated Environmental Monitoring Systems Operations Manual E-5 2. The probe is experiencing “interfer
Turbidity Measurements Appendix E YSI Incorporated Environmental Monitoring Systems Operations Manual E-6
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-26 Set the Comm port number to match the port to which
Flow Appendix F YSI Incorporated Environmental Monitoring Systems Operations Manual F-1 APPENDIX F FLOW Flow is only available in the Sensors
Flow Appendix F YSI Incorporated Environmental Monitoring Systems Operations Manual F-2 Method Type Size Weir V-Notch 22½°, 30°, 45°, 60°, 90°,
Flow Appendix F YSI Incorporated Environmental Monitoring Systems Operations Manual F-3 The final step is to return to the Report menu and choose
Flow Appendix F YSI Incorporated Environmental Monitoring Systems Operations Manual F-4 Deploy the sonde so that the pressure sensor on the side
Using Vented Level Appendix G YSI Incorporated Environmental Monitoring Systems Operations Manual G-1 APPENDIX G USING VENTED LEVEL Sondes that
Using Vented Level Appendix G YSI Incorporated Environmental Monitoring Systems Operations Manual G-2 1. Place the short length of 1/4” tubing
EMC Performance Appendix H YSI Incorporated Environmental Monitoring Systems Operations Manual H-1 APPENDIX H EMC PERFORMANCE When using the YS
EMC Performance Appendix H YSI Incorporated Environmental Monitoring Systems Operations Manual H-2
Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-1 APPENDIX I CHLOROPHYLL MEASUREMEN
Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-2 6025 should approximately track the c
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-27 Figure 39 If your sonde has previously been used, the
Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-3 Note: Be sure to carr
Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-4 Will the chlorophyll values determine
Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-5 What Data Filter settings should I u
Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-6 available water sampling devices that
Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-7 Note also that your may observe less
Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-8 Note the significant improvement i
Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-9 Do I have to worry about the effect o
Chlorophyll Measurements Appendix I YSI Incorporated Environmental Monitoring Systems Operations Manual I-10
Percent Air Saturation Appendix J YSI Incorporated Environmental Monitoring Systems Operations Manual J-1 APPENDIX J PERCENT AIR SATURATION The
Percent Air Saturation Appendix J YSI Incorporated Environmental Monitoring Systems Operations Manual J-2 The “DOsat %Local” Convention In the s
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-28 Figure 40 - Sonde Menu Flow Chart SYSTEM SETUP A
Percent Air Saturation Appendix J YSI Incorporated Environmental Monitoring Systems Operations Manual J-3 %Local” values. For example, assume th
Percent Air Saturation Appendix J YSI Incorporated Environmental Monitoring Systems Operations Manual J-4
PAR Sensor Appendix K YSI Incorporated Environmental Monitoring Systems Operations Manual K-1 APPENDIX K PAR SENSOR Some users of the YSI 6600-
PAR Sensor Appendix K YSI Incorporated Environmental Monitoring Systems Operations Manual K-2 To set up your special PAR 66
PAR Sensor Appendix K YSI Incorporated Environmental Monitoring Systems Operations Manual K-3 Finally, if you want to present your PAR data in ph
PAR Sensor Appendix K YSI Incorporated Environmental Monitoring Systems Operations Manual K-4
Protective Zinc Anode Appendix L YSI Incorporated Environmental Monitoring Systems Operations Manual L-1 APPENDIX L PROTECTIVE ZINC ANODE Shipb
Protective Zinc Anode Appendix L YSI Incorporated Environmental Monitoring Systems Operations Manual L-2 1. Using a large regular blade screwdr
Protective Zinc Anode Appendix L YSI Incorporated Environmental Monitoring Systems Operations Manual L-3
Protective Zinc Anode Appendix L YSI Incorporated Environmental Monitoring Systems Operations Manual L-4 NOTE: The zinc anode collar must be remo
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-29 Select 1-Date & time. An asterisk will appear next
ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-1 APPENDIX M ROX OPTICAL DO SENSOR This
ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-2 3. The ROX sensor is automatically wipe
ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-3 • The 6150 optical sensor has no flow d
ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-4 Should I calibrate my ROX sensor in wate
ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-5 • When prompted for the input of a baro
ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-6 How often should I change my ROX sensor
ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-7 How should I store my ROX sensor when it
ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-8 guard. It is also important to note th
ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-9 protocol will prevent the stiff EDS auxi
ROX Optical DO Sensor Appendix M YSI Incorporated Environmental Monitoring Systems Operations Manual M-10
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-30 ENABLING SENSORS To activate the sensors that are in y
NMEA Applications Appendix N YSI Incorporated Environmental Monitoring Systems Operations Manual N-1 APPENDIX N NMEA APPLICATIONS 6-series so
NMEA Applications Appendix N YSI Incorporated Environmental Monitoring Systems Operations Manual N-2 7, SpCond uS/cm 4, Cond mS/cm 5, Cond
NMEA Applications Appendix N YSI Incorporated Environmental Monitoring Systems Operations Manual N-3 204, Rhodamine ug/L 211, ODOsat % 214, ODOsa
NMEA Applications Appendix N YSI Incorporated Environmental Monitoring Systems Operations Manual N-4
Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-1 APPENDIX O SPECIFICATIONS SONDE SPECIFICATI
Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-2 Temperature: -5 to +50 C for most sensors Dep
Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-3 Computer Interface: RS-232C, SDI-12 Internal
Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-4 6920V2-2 SONDE Available Sensors Temperatur
Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-5 Storage Temperature: -40 to +60 C for sonde
Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-6 600XLM SONDE Available Sensors Temperature, C
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-31 Note that the exact appearance of this menu will vary
Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-7 Internal logging memory size: 384 kilobytes
Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-8 Diameter: 1.65 inches (4.2 cm) Length:
Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-9 600LS SONDE Available Sensors Temperature, Co
Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-10 Non-Vented Level - Shallow Sensor Type...
Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-11 pH Sensor Type...Glass combination ele
Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-12 Rhodamine WT Sensor type... Optical,
Specifications Appendix O YSI Incorporated Environmental Monitoring Systems Operations Manual O-13 Ammonia-Nitrogen Sensor Type...Calcula
Item # 069300 Drawing # A69300 Revision H November 2011 1700/1725 Brannum Lane Yellow Springs, Ohio 45387 USA (800) 765-4974 (937) 767-7241 FAX: (
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-32 CHECKING ADVANCED SETTINGS Select Advanced from the Ma
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-33 If you have a depth sensor installed, you can maintai
5.14 CHLOROPHYLL 5-21 5.15 RHODAMINE 5-28 5.16 PHYCOCYANIN-CONTAINING BLUE-GREEN ALGAE 5-31 5.17 PHYCOERYTH
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-34 2.6 GETTING READY TO CALIBRATE 2.6.1 INTRODUCTION
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-35 6. Have several clean, absorbent paper towels or co
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-36 RECOMMENDED VOLUMES OF CALIBRATION REAGENTS The approx
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-37 Table 6 6600V2-4 Sonde with Standard Long Calibrati
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-38 NOTE CAREFULLY: All optical sensors MUST be calibrate
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-39 3. Note that the exact appearance of this menu will va
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-40 Note that all parameters that have been enabled will
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-41 NOTE: The YSI conductivity system is very linear over i
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-42 Monitoring Applications If your instrument will be use
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-43 DEPTH AND LEVEL For the depth and level calibration, m
Introduction Section 1 YSI Incorporated Environmental Monitoring Systems Manual 1-1 SECTION 1 INTRODUCTION 1.1 ABOUT YSI INCORPORATED From a
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-44 The next calibration instructions are only for the ISE
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-45 CALIBRATION TIP: Exposure to the high ionic content
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-46 water. Input the value 0 NTU at the prompt, and press
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-47 Before making any field readings, carefully read Sectio
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-48 RHODAMINE WT 2-POINT Select Rhodamine from the Calibra
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-49 2.7 TAKING READINGS After you have (1) enabled the se
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-50 If you started sampling without entering a filename, th
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-51 sampling at 6:00 PM on July 31, 1996. The site is Clea
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-52 Skip B-View params to log in this initial test stud
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-53 The display now shows the next date and time for loggi
Introduction Section 1 YSI Incorporated Environmental Monitoring Systems Manual 1-2 techniques, and system setup is necessary to obtain accurate
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-54 2.8 USING ECOWATCH TO CAPTURE, UPLOAD AND ANALYZE DAT
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-55 Select 1-Directory to view all fil
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-56 Select 1-Proceed. Choose the appropriate file transfe
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-57 Select 5-Quick view file to view the last page of data
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-58 Figure 41 Opening a File Some daily variations may b
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-59 Figure 42 Viewing Options It may be somewhat awkward
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-60 Figure 43 Viewing the Data in Table Format Viewing
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-61 Figure 44 Statistics and Study Information Next,
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-62 Figure 45 Viewing the Data with Right-Button Mouse Fun
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-63 Figure 46 Changing the Appearance of a Graph or Tabl
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-1 SECTION 2 SONDES 2.1 GETTING STARTED The 6-Series En
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-64 along the x-axis, you may also manually scale the y-axi
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-65 Figure 48 Using Limit Data Set to Display a Subset of
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-66 Figure 49 Saving, Exporting, Printing and Related Func
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-67 Figure 50 Customizing a Graph from SAMPLE.DAT Final
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-68 2.9 SONDE MENU The functions of the sondes a
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-69 2.9.1 RUN Select 1-Run from the Main menu to begin t
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-70 Select 1 – Start sampling option to start discrete samp
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-71 Sampling Faster Than 0.5 Seconds For special applicati
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-72 Follow the prompts on this screen to prepare your sond
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-73 Select B-View Parameters to log to confirm that your
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-2 2.2 CONNECTING YOUR SONDE There are a number of ways i
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-74 Select 1-Yes and the screen will change. The displa
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-75 2.9.2 CALIBRATE All of the sonde sensors (except temp
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-76 CONDUCTIVITY Select 1-Conductivity to calibrate the c
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-77 For the mg/L mode, calibration is carried out in a wate
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-78 membrane installed on the 6150 probe and NOT a function
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-79 which characterize the new membrane and they MUST be en
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-80 Calibration Using mg/L – 1-Point Place the sensor in a
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-81 NOTE CAREFULLY: If you used sodium sulfite solution as
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-82 pH When selecting ISE1 pH, you will be given the choic
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-83 When selecting ISE3-NH4+, you will be given the choice
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-3 Figure 2 Sonde to Data Collection PlatformYou will need
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-84 CHLORIDE When selecting ISE5-CL-, you will be given th
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-85 When selecting Optic X -6026-Turbidity (or 6136-Turbidi
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-86 Enter after the readings have stabilized to confirm the
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-87 To calibrate your 6131 Blue-Green Algae-Phycocyanin (BG
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-88 determining the BGA-PC content of algal suspensions by
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-89 point calibration, but the software will prompt you to
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-90 CALIBRATION RECORD – THE GLP FILE When any sensor is c
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-91 Pressing 1-View file will show the calibration record f
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-92 Constants and is described in Section 2.9.8. All other
Sondes Section 2 YSI Incorporated Environmental Monitoring Systems Operations Manual 2-93 PC6000 format will transfer the data so that it wil