Página 1 - Operating manual

ba75793e02 12/2008pH 3210Operating manualpH meterHOLD AR25.06.093pH°CMenu15.11.200808:37

Página 2 - © Weilheim 2008, WTW GmbH

Safety pH 321010ba75793e02 12/20082.1 Authorized useThis meter is authorized exclusively for pH and ORP measurements in a laboratory or field environm

Página 3

pH 3210 Safety11ba75793e02 12/2008Safe operation If safe operation is no longer possible, the meter must be taken out of service and secured against i

Página 4 - Contents pH 3210

Safety pH 321012ba75793e02 12/2008

Página 5 - 1Overview

pH 3210 Commissioning13ba75793e02 12/20083 Commissioning3.1 Scope of deliveryz pH meter, pH 3210 z 4 batteries 1.5 V Mignon type AAz Short instruction

Página 6 - 1.1 Keypad

Commissioning pH 321014ba75793e02 12/2008NoteAlternatively, you can also use Ni-MH rechargeable batteries (type Mignon AA). In order to charge the bat

Página 7 - 1.2 Display

pH 3210 Operation15ba75793e02 12/20084 Operation4.1 Switching on the meterSwitching on Press the <On/Off> key.The meter performs a self-test. Th

Página 8 - 1.3 Socket field

Operation pH 321016ba75793e02 12/20084.2 General operating principlesThis section contains basic information on the operation of the pH 3210. Operatin

Página 9 - 2 Safety

pH 3210 Operation17ba75793e02 12/20084.2.2 NavigationMeasured value display In the measured value display, you can z open the respective measuring men

Página 10 - ECHNICAL DATA

Operation pH 321018ba75793e02 12/2008z FunctionsFunctions are designated by the name of the function. They are immediately carried out by confirming w

Página 11

pH 3210 Operation19ba75793e02 12/20084.2.3 Example 1 on navigation: Setting the language1 Press the <On/Off> key.The measured value display appe

Página 12 - Safety pH 3210

pH 3210ba75793e02 12/2008Accuracy whengoing to pressThe use of advanced technology and the high quality standard of our instruments are the result of

Página 13 - 3 Commissioning

Operation pH 321020ba75793e02 12/20085 Select the General submenu with <▲><▼>. The current selection is displayed with a frame.6 Open the

Página 14 - Commissioning pH 3210

pH 3210 Operation21ba75793e02 12/20089 Confirm the setting with <ENTER>. The meter switches to the measuring mode.The selected language is activ

Página 15 - 4 Operation

Operation pH 321022ba75793e02 12/20084.2.4 Example 2 on navigation: Setting the date and timeThe meter has a clock with a date function. The date and

Página 16 - Operation pH 3210

pH 3210 Operation23ba75793e02 12/20084 Change and confirm the setting with <▲><▼> and <ENTER>.The minutes are highlighted.5 Change a

Página 17

Operation pH 321024ba75793e02 12/20084.3 Sensor-independent settingsThe Storage & config menu comprises the following settings:z System (see secti

Página 18

pH 3210 Operation25ba75793e02 12/20084.3.2 Data storageThis menu contains all functions to display, edit and erase stored measured values and calibrat

Página 19 - Data storage

Operation pH 321026ba75793e02 12/20084.3.3 Automatic Stability controlThe automatic Stability control function continuously checks the stability of th

Página 20

pH 3210 Operation27ba75793e02 12/20084.4 pH value / ORP voltage4.4.1 General informationYou can measure the following variables:z pH value [ ]z ORP [m

Página 21

Operation pH 321028ba75793e02 12/20084.4.2 Measuring the pH valueStability control(AutoRead )The stability control function (AutoRead) continually che

Página 22 - 14:53:40

pH 3210 Operation29ba75793e02 12/2008Criteria for a stablemeasured valueThe Stability control function checks whether the measured values are stable w

Página 23

pH 3210 Contents3ba75793d02 12/2008pH 3210 - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.1 Keypa

Página 24 - _>/[Menu] or

Operation pH 321030ba75793e02 12/20084.4.3 Measuring the ORPNoteORP electrodes are not calibrated. However, you can check ORP electrodes using a test

Página 25

pH 3210 Operation31ba75793e02 12/2008prematurely terminated, the current measurement data are not output to the interface.Criteria for a stablemeasure

Página 26

Operation pH 321032ba75793e02 12/20084.4.4 Settings for pH and ORP measurementsOverview The following settings are possible for pH and ORP measurement

Página 27 - 4.4 pH value / ORP voltage

pH 3210 Operation33ba75793e02 12/2008Temperature unit °C °FTemperature unit, degrees Celsius or degrees Fahrenheit.All temperatures are displayed with

Página 28

Operation pH 321034ba75793e02 12/20084.4.5 pH calibrationWhy calibrate? pH combination electrodes age. This changes the zero point (asymmetry) and slo

Página 29 - Δ pH: better than 0.01

pH 3210 Operation35ba75793e02 12/20087 Merck 4 * 2.0004.0007.00010.00020°C8 Merck 5* 4.0107.00010.00025 °C9 DIN 19267 1.0904.6506.7909.23025 °C10 Mett

Página 30 - 157.0 mV

Operation pH 321036ba75793e02 12/2008* Brand names or trade names are trademarks of their respective owners protected by law.NoteThe buffers are selec

Página 31 - Δ mV: better than 0.3

pH 3210 Operation37ba75793e02 12/2008Calibration points Calibration can be performed using one to five buffer solutions in any order (single-point to

Página 32

Operation pH 321038ba75793e02 12/2008+ -25 ... +25 -61 ... -60.5 or -57 ... -56- -30 ... +30 -62 ... -61 or -56 ... -50Clean the combination electrode

Página 33

pH 3210 Operation39ba75793e02 12/2008Preparatory activities Perform the following preparatory activities when you want to calibrate:4.4.6 Calibration

Página 34

Contents pH 32104ba75793d02 12/20084.5 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 494.5.1 Manual storage . .

Página 35

Operation pH 321040ba75793e02 12/20084.4.7 Carrying out an automatic calibration (AutoCal)Make sure that the correct buffer set is selected in the sen

Página 36

pH 3210 Operation41ba75793e02 12/2008NoteFor single-point calibration, the instrument uses the Nernst slope (-59.16 mV/pH at 25 °C) and determines the

Página 37

Operation pH 321042ba75793e02 12/2008Continuing with three-to five-point calibration(Buffer TEC)12 Wait for the measurement with stability control to

Página 38 - HAT TO DO IF... (page 57)

pH 3210 Operation43ba75793e02 12/2008NoteCalibration is automatically completed after the last buffer of a buffer set has been measured. Then the cali

Página 39

Operation pH 321044ba75793e02 12/20084.4.8 Carrying out a manual calibration (ConCal)Single-point calibration Use any buffer solution for this rapid m

Página 40

pH 3210 Operation45ba75793e02 12/2008NoteFor single-point calibration, the instrument uses the Nernst slope (-59.16 mV/pH at 25 °C) and determines the

Página 41

Operation pH 321046ba75793e02 12/2008Continuing with two-point calibration10 Thoroughly rinse the combination electrode with distilled water.11 Immers

Página 42

pH 3210 Operation47ba75793e02 12/20084.4.9 Displaying calibration recordsDisplaying thecalibration recordThe calibration record of the last calibratio

Página 43 - 22.09.2008

Operation pH 321048ba75793e02 12/20084.4.10 Continuous measurement control (CMC function)The continuous measurement control (CMC function) enables to

Página 44

pH 3210 Operation49ba75793e02 12/20084.5 StorageYou can transmit measured values (datasets) to the data storage:z Manual storage (see section 4.5.1)Me

Página 45

pH 3210 Overview5ba75793e02 12/20081OverviewThe compact pH 3210 precision pH meter enables you to perform pH measurements rapidly and reliably. The pH

Página 46

Operation pH 321050ba75793e02 12/20084.5.1 Manual storageYou can transmit a measurement dataset to the data storage as follows: If the storage is full

Página 47 - _> key

pH 3210 Operation51ba75793e02 12/20084.5.2 Editing the measured value storageThe contents of the manual or automatic measurement data storage can be s

Página 48

Operation pH 321052ba75793e02 12/2008following options:z Switch directly to the measured value display with <M>.z Quit the display and move to t

Página 49 - 4.5 Storage

pH 3210 Operation53ba75793e02 12/20084.6 ResetYou can reset (initialize) all sensor settings and sensor-independent settings separately from each othe

Página 50

Operation pH 321054ba75793e02 12/20084.6.2 Resetting the system settingsThe following system settings can be reset to the delivery status:The system s

Página 51 - _>/[Menu] key in the

pH 3210 Maintenance, cleaning, disposal55ba75793e02 12/20085 Maintenance, cleaning, disposal5.1 MaintenanceThe only maintenance activity required is r

Página 52

Maintenance, cleaning, disposal pH 321056ba75793e02 12/2008battery compartment. 5.2 CleaningOccasionally wipe the outside of the meter with a damp, li

Página 53 - 4.6 Reset

pH 3210 What to do if...57ba75793e02 12/20086 What to do if...Error messageOFL, UFLError messageErrorCause RemedypH combination electrode:– Measured v

Página 54 - _>/[Menu]

What to do if... pH 321058ba75793e02 12/2008No stable measuredvalue Sensor symbol flashesDisplayCause RemedypH combination electrode:– Junction contam

Página 55 - 5.1 Maintenance

pH 3210 What to do if...59ba75793e02 12/2008Obviously incorrectmeasured valuesMeter does not react tokeystrokeYou want to know whichsoftwareversion is

Página 56 - 5.4 Disposal

Overview pH 32106ba75793e02 12/20081.1 KeypadIn this operating manual, keys are indicated by brackets <..> . The key symbol (e.g. <ENTER>)

Página 57 - 6 What to do if

What to do if... pH 321060ba75793e02 12/2008

Página 58 - AINTENANCE)

pH 3210 Technical data61ba75793e02 12/20087 Technical data 7.1 General dataDimensions approx. 180 x 80 x 55 mmWeight approx. 0.4 kgMechanical structur

Página 59 - _>[Menu] / Storage &

Technical data pH 321062ba75793e02 12/20087.2 Measuring ranges, resolution, accuracyNoteThe accuracy values specified here apply exclusively to the me

Página 60 - What to do if... pH 3210

pH 3210 Technical data63ba75793e02 12/2008FCC Class A Equipment StatementNote: This equipment has been tested and found to comply with the limits for

Página 61 - 7 Technical data

Technical data pH 321064ba75793e02 12/2008

Página 62 - Technical data pH 3210

pH 3210 Lists65ba75793e02 12/20088Lists This chapter provides additional information and orientation aids.Specialist terms The glossary briefly explai

Página 63

Lists pH 321066ba75793e02 12/2008Molality Molality is the quantity (in Mol) of a dissolved substance in 1000 g solvent.ORP voltage The ORP is caused b

Página 64

pH 3210 ListsIndex67ba75793d02 12/2008AAuthorized use . . . . . . . . . . . . . . . . . . . . . . 10Automatic switch-off . . . . . . . . . . . . . . .

Página 65

Lists pH 321068ba75793d02 12/2008

Página 67

pH 3210 Overview7ba75793e02 12/20081.2 DisplayFunction displayindicators<ENTER>: <ENTER_>:Opens the menu for measurement settings / confir

Página 68

Wissenschaftlich-Technische Werkstätten GmbHDr.-Karl-Slevogt-Straße 1D-82362 WeilheimGermanyTel: +49 (0) 881 183-0+49 (0) 881 183-100Fax: +49 (0) 881

Página 69

Overview pH 32108ba75793e02 12/20081.3 Socket fieldCautionOnly connect sensors to the meter that cannot return any voltages or currents that are not a

Página 70

pH 3210 Safety9ba75793e02 12/20082 SafetyThis operating manual contains basic instructions that you must follow during the commissioning, operation an