Click on the Low EU field and enter the
low engineering unit that will be used
to calibrate the logger, not necessarily
the lowest reading the sensor can output.
In this example, the value is zero feet,
which
is
the lowest calibration point and water level.
Click Next.
Place the sensor in the condition that will
cause it to output the level corresponding
to the low EU entered in the previous menu.
Wait for the data reading to stabilize.
Click Next to complete the calibration.